General Warnings and Cautions
SD32+
System
4
Immunity Test
IEC 60601-1-2 Test Level
Compliance
Electromagnetic
environment
Magnetic fields
at ground
frequency
(50/60Hz)
IEC 61000-4-8
3 A/m
Radiated RF
fields
IEC 61000-4-3
Non-life-supporting
equipment
3 V/m
From 80 MHz to 2.5 GHz
IEC 60601-1-2 Test
Levels
Residential
(Note 4)
Conducted RF
fields
IEC 61000-4-6
Non-life-supporting
equipment
3 V
From 150 kHz to 80 MHz
Measures to observe
Note 1 – The floor should be antistatic (wood, ceramic, etc.). If carpeted, the relative humidity
should be maintained at 30%.
Note 2 – The main power supply must be of residential, commercial, or medical grade.
Note 3 – If operating for an extended period, a UPS system must be provided in the event of a
power outage.
Note 4 – Mobile or portable RF communication devices should be used at greater distances than
those provided in table 3.
Table 2. electromagnetic immunity
1.5.1. recommended distance from Radio Frequency systems
It is not recommended to use RF transmitting devices near the
SD32+
digital amplifier. RF
transmitting devices can cause interference with the device and impact the quality of the signals
being acquired during use.
To prevent RF interference it is recommended to maintain a minimum distance between the
SD32+
digital amplifier and the RF transmitting device being used; such as cell/mobile phones.
The following table shows recommended distances that the
SD32+
digital amplifier should
remain from some common RF transmitting devices.
RF Source
Typical Rated
Power (W)
Distance
(m)
Microcellular phone CT1, CT2, CT3
0.01
0.3
DECT cellular phone, Wireless Information
Technology equipment (modems, LAN's)
0.25
2
Cellular phone, hand-held (USA)
0.6
2
Cellular phone, hand-held (GSM and NMT,
Europe DECS 1800)
2
8
4
7
Walkie-talkie (Rescue, police, fire, maintenance)
5
3
Cellular phone, bag
16
10
Mobile radio (Rescue, police, fire)
100
30