EM TEST
VDS 200 Series
Manual for Operation
V 5.19
20 / 57
3.3.2. ISO 16750-2 WD 03/2000-2
3.3.2.1. Short voltage drop
This test is to simulate the effect of a classical fuse actuation in another circuit.
tr, tf =
10ms
Limits depends VDS voltage range
Input restrictions
0.0 V
Vb + Va1
30.0V (60.0V)
0.0 V
Vb + Va2
30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t2
0.1s
-
99.9s
t3
0.1s
-
99.9s
int
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
30A (Imax)
3.3.2.2. Slow decrease / increase
This test is to simulate a gradual discharge and recharge of the battery.
Parameters:
V1
0.0V
-
30.0V (60.0V)
V2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t2
0.1s
-
9999.9s
I
1A
-
30A (Imax)
Remarks:
WD 03/2000-2 Voltage change rate = ( 3
0.1 ) V per minute
ISO 16750-2
Voltage change rate = ( 0.5
0.1 ) V per minute