6.3.16 Continuous operation
Continuous operation can be established to produce a repetitive stream of discharges.
Pushing the trigger button starts the continuous operation; pressing it again stops the operation.
WARNING
Continuous operation should only be utilized in cases of real necessity since every ESD radiates electro-
magnetic disturbance the effect of which on the environment must be taken into consideration. The test area
should be made out of bounds for unauthorized personnel. The test must be monitored throughout its dura-
tion. When running on battery the duration of the test is naturally limited.
6.4
The Home Screen
The home screen allows three methods of operation:
•
Quick Start
– Basic operation whereby users may set the voltage, choose test levels, air or contact discharge and
timings.
•
Standards
– Where the standards are found for users who would like to go straight to the standard testing with lim-
ited ability to adjust settings.
•
Test Routing
– Select a previously stored test routine that has been sent from esd.control
6.4.1
Quick Start
This is fastest way to get started and all modes, voltage, discharge networks etc. can be used immediately.
The
6.4.2
Standards
Standards are pre-programmed for most IEC and ISO testing. The levels are pre-programmed for air and contact discharge
and, in the case of automotive, the R/C network and the category as well.
In Standards mode, when the test level is selected, the type of tip (air or contact discharge) is taken into account, as well as
verifying that the proper R/C network is used.
6.4.3
Test Levels
In addition to all basic settings found in this guide, users can also set various test levels for use during Quick Start mode, or
may turn them off completely.
To set the test levels, choose Test Level 1 through 4 (TL1..TL4) and enter the voltage for both contact discharge (CD) and
air discharge (AD) and tap “ON” and confirm with “OK”. Conversely, if you do not wish to use test levels, set this to “OFF”.