6.3.13 Threshold
This function permits different sensitivity levels to be set for the arcing detector whereby a differentiation can be made be-
tween stray discharges and a true discharge onto the EUT.
Normal
Arcing is detected and is indicated by the kV symbol on the display blinking provided 20% (or more) of the charge voltage is
dissipated.
Low
Arcing is detected and is indicated by the kV symbol on the display blinking provided 10% (or more) of the charge voltage is
dissipated.
High
Arcing is detected and is indicated by the kV symbol on the display blinking provided 30% (or more) of the charge voltage is
dissipated.
Off
This position is made for EUTs with non-conductive surfaces (housings) and counts every contact discharge, and every time
the air discharge is stopped with the trigger.
6.3.14 Display Settings
The Display Settings menu allows users to set the brightness and to determine how long before the simulator sleeps in bat-
tery mode (esd NX30 only).
6.3.15 ISO-Selftest
Provides the means to select the required, pre-programmed standard test (e.g. IEC / EN 61000-4-2, level 4).
Just starting the ISO-Selftest gives a quick response of the proper operation of the ESD simulator The screen reflects all
required voltage levels as well as the tolerances in table form, based on the ISO 10605 standard.
During the calibration procedure the HV module is strained up to 30 kV in steps, and the charge voltage is measured,
checked if in tolerances and validated for each step.
If somehow the maximum voltage could not be reached or hold, the esd NX30 comes out with an error message Nr. 210 cal.
Failed. The calibration procedure is diagnosing, to which voltage level the instrument works properly. This voltage value will
be shown on screen. The instrument can be used for tests up to this voltage value, but it is recommended to return it to a
EM Test/AMETEK service center for repair.