EM TEST
EFT 500 Series
Manual for Operation
V 2.00
9 / 21
2.5.
Test Routine as per IEC 61000-4-4 (2004-07) Ed 2
The display shows a list of test levels as per the standards IEC 61000-4-4. In the menu Service F4Change
standard level you select between the burst frequency 5kHz and 100kHz.
Setting IEC 61000-4-4
Setting IEC 61000-4-4 Ed 2
STANDARD TEST ROUTINES
STANDARD TEST ROUTINES
F1
:
Level 1
260 V /
5.0 kHz
F1
:
Level 1
260 V /
100.0 kHz
F2
:
Level 2
500 V /
5.0 kHz
F2
:
Level 2
500 V /
100.0 kHz
F3
:
Level 3
1000 V /
5.0 kHz
F3
:
Level 3
1000 V /
100.0 kHz
F4
:
Level 4
2000 V /
5.0 kHz
F4
:
Level 4
2000 V /
100.0 kHz
F5
:
Level 5
4000 V /
5.0 kHz
F5
:
Level 5
4000 V /
100.0 kHz
F6
:
Level X - Level Y
F6
:
Level X - Level Y
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
The function key F5 selects a procedure which starts at test level X and stops at test level Y. The test level is
changed automatically after the preselected test time T.
(X <= Y)
Page 3
(Show values)
Standard IEC 1000-4-4
Level 3
V
=
2000V
f
:
100 kHz
td
=
0.8ms
tr
:
300ms
kop
=
L N PE
+/- :
+
T
=
1:00min
The Standard defines the burst duration at
100kHz with 0.75ms. The EFT 500 does not
support a resolution of 0.05ms.
The burst duration at 100kHz is set to 0.8ms
which is inside the specified tolerance of the
standard of 0.75ms
±
20%
F1
F2
F3
F4
F5
F6
The functions START, CHANGE and PRINT are the same as defined under Quick Start.
The function key CHANGE can only handle the coupling, polarity and the test time.
All other parameters are defined by the standard.
2.6.
Service
SERVICE
F1 : Address EM TEST AG
F2 : Selftest
F3 : Set-up
F4 : Change standard levels
F5 : Print all
F1
F2
F3
F4
F5
F6
F7
F1 Addresses
All important data regarding EM TEST are indicated.
F2 Selftest
Selftest for check the key functions.
F3 Set-up
The software will clearly explain the set-up procedure.
F4 Change standard levels
This procedure allows the user to change the standard values to his actual requirements.
F5 Print all
All stored settings are printed to the serial interface RS 232.