EM Test
CWS500 N2
User manual
V 3.00
32 / 38
7.2.
CWS500N2.1
Summary:
Output power
150W (nominal)
Output impedance
50 Ohm, VSWR 2.0: 1 nominal
Level approx.
-13dBm ... 51.7dBm
Current 4MHz ... 400MHz
max. 950mA (150W amplifier, without attenuator, F-130A clamp)
Test frequencies ( synthesizer )
Sinusoidal (CW)
9kHz - 1000MHz
Frequency bands
9kHz - 1000MHz
step: 1kHz ( icd.control and Quickstart online)
Quickstart setting
9kHz - 9,999MHz
step: 1kHz
10MHz - 99,99MHz
step: 9kHz
100MHz - 1000MHz
step: 100kHz
Modulation
AM
Frequency:
1Hz ... 3000Hz
Index:
0%...95%
PM
Frequency:
1Hz ... 3000Hz
Duty cycle:
10%...80%
Attenuator
Range
0 dB - -63.5dB
Step level
0.5dB
Timing:
Dwell time
general
td = 0,3s - 9999.9s
pulse modulation
td = 3s - 9999.9s
Rest time
tr = 0 / 0,3s - 9999.9s
Output
Direct
N type connector
EUT control
BNC input FAIL 1
Fail 1; test will be stopped immediately
BNC input FAIL 2
Fail 2; failure is detected, test continues
Routines (integrated)
Service
Service, Setup
Test routines
All test only in remote control with icd.control software version >3.00