Pulse count (Pull up)
–
S0 input
S0 signal between input and ground.
Select “Pulse input
(pull up)” in
the
application “Sensor Settings”. ELT input is internally
pulled up by ~50 k
Ω
*.
4
–
20 mA input
Select “Analog input” in
the application
“Sensor Settings”. Connect resistor parallel to the
input. Value of R can be calculated as
𝑈 = 𝑅 ∙ 𝐼
. For full range, use 500
Ω
resistor.
(20mA will then read as 10V).