Temperature-Volume Corrector TC210
18
ELSTER Handel GmbH
2.4 Changing values
2.4.1 Differentiating between values (data classes)
The methods of entering and changing values differ depending on the value. These are
therefore subdivided into so-called "data classes" (abbreviation: "DC"). Values in the same
data class are treated identically during entry. A prerequisite for an entry is that the lock
assigned to the value is open.
The following data classes (DC) are present in the TC210:
DC
Type
Entry, change using "ENTER"
1
Display test
No change possible.
2
Function
Function is initiated by pressing the key for at least two seconds.
3
Constant
No change possible.
4
Measurement
The value is updated by pressing the key for at least two seconds.
5
Status
No change possible.
7
Discrete value
When the key is pressed for at least two seconds, the value can
be changed by using the key to select from a list of possible val-
ues.
8
Permanent
value
No change is possible using the keypad on the device.
11
Combination
Changes possible on the device, but with masked entry, i.e. only
the character currently being edited is visible, all others are
masked out by a minus sign.
With a closed lock it is opened on entering the correct combina-
tion.
With an open lock, the combination is changed by the entry.
12
Counters
As "Permanent value" (see above.).
15
Computation
counter
No change possible.
16
Initial value
No change possible, sometimes branching to a submenu.
19
Status register
No change possible.
If a value is accommodated in a submenu, it cannot be changed independent of its data
class by the keypad, since pressing the key for at least two minutes then results in branch-
ing into the submenu.
Содержание TC210
Страница 2: ...Temperature Volume Corrector TC210 2 ELSTER Handel GmbH ...
Страница 8: ...Temperature Volume Corrector TC210 8 ELSTER Handel GmbH ...
Страница 51: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 51 ...
Страница 52: ...Temperature Volume Corrector TC210 52 ELSTER Handel GmbH ...
Страница 53: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 53 Part 2 Description of the Initial Operation ...
Страница 62: ...Temperature Volume Corrector TC210 62 ELSTER Handel GmbH ...
Страница 63: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 63 Appendix ...
Страница 66: ...Temperature Volume Corrector TC210 66 ELSTER Handel GmbH A 2 Certificate for Ex Zone 1 ...
Страница 67: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 67 ...
Страница 68: ...Temperature Volume Corrector TC210 68 ELSTER Handel GmbH ...