Elster NV/SA
Rijkmakerlaan 9
2910 Essen,
Belgium
T + 32 3 670 0700
www.elster-instromet.com
DOCUMENT TITLE
ULTRASONIC FLOWMETER SERIES 6, MODELS Q.SONIC
PLUS
SUBTITLE
QUICK START MANUAL
DOCUMENT NUMBER
22.100.200.000.02/2
REVISION
B
REVISION DATE
2013-02-22
This reproduction is sole property of Elster NV/SA and is subject to the conditions that it or any information contained therein will not be used in any way
detrimental to our interests and that all copies will be returned immediately on demand
PAGE
19 OF 29
22.100.200.000.02_2_B ULTRASONIC FLOWMETER SERIES 6.DOC
5.6
Hot commissioning
The hot commissioning is in most cases the last test of the UFM and can be witnessed by a
representative of the client and, if required, by an inspector of weights and measures for the
official sealing. Under this condition there is process gas in the pipe and a flow test is being
performed. The performance, AGC Levels/Limits and zero flow are checked again. If
possible, the gas flow running through the UFM will be compared to another flowmeter in
the line. Most UFM are calibrated gas flow meters, so the measured value is reliable
without exception. Furthermore, the interaction with the flow computer can be tested.