![ELPRO ECOLOG-PRO Series Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/elpro/ecolog-pro-series/ecolog-pro-series_operation-manual_2395990010.webp)
Document: EP6001Eb
ECOLOG-PRO 4PT PT100 MEASUREMENT MODULE
10 - EN
3
ECOLOG-PRO 4PT Pt100 Measurement Module
Module
Measurement module for up to 4 Pt100 sensors
• Circuit: 4-wire connection
• Sensor class: A, recommended
• Sensor details
FP2005E.
Status indicator 1
• green; power supply ok
• blue; system bus activity
Status indicator 2
ECOLOG-PRO 4PT, 801420
Measurement range
-200°C to 200°C
Measurement accuracy
of module
±0.2°C
Sensor error and cable fault are not considered.
Measurement resolution
±0.1°C
Cable fault
-
up to 30 m 4x0.25 mm
2
without
shield.
-
30 m to 50 m 4x0.25 mm
2
with
shield.
-
50 m to 80 m 4x0.34 mm
2
with
shield.
±0.06°C
±0.1°C
±0.2°C
Logging interval
1 minute to 60 minutes
Each channel is capable of logging at a separate
interval.
Date and time
Internal system clock
The clock is synchronized permanently to elpro-
MONITOR system time.
Module power supply
Via internal system bus
Data security
-
Internal non-volatile memory
-
10,000 measurements
In case the network connection fails, measure-
ments continue to be logged and stored in the
internal memory.