Replace the Nasal Cannula with a new one if there is a warning on LCD screen or once in every 30
days.
CLEANING AND CHANGING OF WATER BOWEL:
In every 7 days clean with soap and water and rinse properly and replace with a new one when
there will be a warning on LCD screen or once in every 90 days.
6.
TROUBLSHOOTING:
OBSERVATION
PROBABLE REASON/CAUSES
SOLUTION
On opening there is no
alarm and device is not
running.
Electrical power supply cord is not
properly connected either with device or
in plug.
There is no electricity in the plug, fuse
burnt out.
Connect the electrical cord properly
both ends. Test if there is electric in
the plug or put the cord in another
plug.
Change the fuse.
Alarm due to high
temperature.
System may be turned to high
temperature protecting mode.
Clean the black filter located at the
rear side of the device. Check whether
the ventilation is sufficient all around
the device. Cooling fan may not
working. Call authorized technical
service. Keeps the device waiting for
40-50 mins. in switched off position
then reopen, if problem continue call
technical service.
High pressure alarm.
System may turn to high pressure
protecting mode.
Control the Nasal Cannula against
possible breaking, twisting. Control
the water bowel against clogging.
Check whether flow meter is open or
closed. If the situations still continue
please call Authorized Technical
Service.
Low Pressure Alarm.
System may turn to low pressure
protecting mode.
Control the dust filter, if require clean
it. Flow meter may setted at 5 lt./min.
reset it to a lower value. Please check,
intermediate connecting hose may
not be connected. If the problems still
continue call technical service.
Low alarm sound.
Low alarm battery.
Consult with authorized technical
service centre.
Call technical service if the suggested troubleshooters are not helpful for solution of problems.