1.
GENERAL INFORMATION:
This user guide has been prepared to inform about HEALTHTIME / HEALTHTIME PLUS Oxygen
Concentrator to the consumers. Please be sure that you have gone thoroughly this manual before
using the device.
These days a lot of people suffering from cardiac, lung & other respiratory malfunctioning owing
to day to day increase in air pollution level. Large numbers of them are taking the advantage of
concentrated oxygen treatment in hospitals and home as well.
HEALTHTIME / HEALTHTIME PLUS Oxygen Concentrator is designed to supply supplemental
oxygen to persons requiring low flow oxygen therapy at home. The device produces oxygen enriched
products by concentrating the oxygen content of ambient air having oxygen content of 21% using
Pressure Swing Absorption (PSA) technique of NASA to a oxygen purity of 90% (+/- 3%). Hence, it
helps you to receive your most intended oxygen therapy. With the help of nasal cannula linked with
the device the patient can receive high concentrated oxygen directly.
The HEALTHTIME / HEALTHTIME PLUS is easy to use. It has a Safety and Solid body, 2x16 LCD
screen, digital timer to show total using time and outlet oxygen percentage (for OSD device only).
Electrical cut-off warning for safety against high pressure, low pressure & high temperature. Thermo
protected compressor. Light in weight and easy to carry. Audio visual warning (alarm) system for
periodic change of Nasal Cannula, Pull Filter & Water Flask.
2.
SAFETY GUIDELINES:
The HEALTHTIME / HEALTHTIME PLUS must only be used for purposes mentioned in
this manual!
Do not use this device as Life Supporting or Life Sustaining unit. Use your device
according to your doctor’s prescription and advices only, otherwise it may present a risk
to the health of you (patient).
Use of highly concentrated oxygen could be dangerous if not prescribed by authorized
person. Please keep spare oxygen sources such as oxygen tube or bag ready for the
patient considering possible malfunctioning of the device may occur anytime. Keep the
device up right position during transportation to avoid damage in the cabinet. Switch off
the device when not used by the patient.
During transportation or when not in use unplug the power supply cord. Do not allow
unauthorized person to open the cabinet of the device. If the electric supply line suffers
from frequent voltage fluctuation, we suggest using Voltage Regulator for the device to
avoid unwanted damages in the sensitive electronic parts of the device.
Do not unplug the power cable by hanging it on the device or by pulling the cable.
Do not leak water on the device or emerge the device into water. If you think there is a
chance of water may come in contact with the device, please unplug the power cable and
inform technical service.
Do not open the cabinet of the device when power cable connected. Only authorized
technical service personnel is permitted to open the device. There are no parts inside the
device which you can repair.