Elmasonic Med
8.3
Operating modes
Operating
mode
Brief description
Display
Switched off
The unit is being supplied with power and
is switched off.
Off
Standby
The device is switched on. The display is
dimmed after a longer period of inactivity.
The automatic safety shut-off is activated
after 8 hours of inactivity.
On
Setup
The unit is switched off. The
Setup
menu
is activated to change the unit functions.
•
Setup
symbol
• Number of the selected unit function
(parameter)
• Current setting (parameter value)
Normal
mode
The unit is ready to use. Ultrasonic
cleaning can be conducted using
adjustable operating parameters.
• Set operating parameters
During normal mode:
• Current operating status
Programme
mode
The unit is ready to use. 1 of 4 stored
programmes can be selected for ultrasonic
cleaning. The programmes can be
specified and saved.
• Available programmes
• Operating parameters of the selected
programme
No programmes have been saved:
• Programme number 1 and "---" for
duration and temperature
During programme operation:
• Current operating status
9
Set up and connect the unit
CAUTION
Heavy loads (applies to unit sizes Elmasonic Med 500/900)!
Serious injuries to the spine or back muscles.
Ø
Comply with the health and safety requirements on lifting heavy loads.
9.1
Notes on installation site
The unit is designed to be set up in commercially used premises such as laboratories, medical practices,
etc.
• The installation site must be well ventilated.
• The permissible ambient conditions must be correct.
• The unit must be installed on an even, non-slip, moisture-resistant work surface with the following
characteristics: