SwemaFlow 125D
side 14
Technical data
Measuring range:
Air Flow:
2... 125 l/s, 7... 450 m3/h, 4... 260 cfm
Temperature:
0...+50°C, 32...122°F
Barometer:
600... 1200 hPa, 18...35 inHg
Accuracy (95% coverage probability in non condensing, non moist air, <80% RH):
Air Flow:
±3,5% read value, min 0,4 l/s
Using Back-pressure method:
±10% min 1 l/s
Temperature:
±0,5°C
Barometer:
±3,5 hPa
Measuring method:
Mass flow, net of hot wires
Weight:
1,975 kg (including capture hood 300x300mm and throttle ring )