background image

 Elma 3037/3039

 

15 

Instrument 

CAT III 600V with respect to earuth for the jaw 

      The barrier on the jaw is indicating the limit of 
safe access of the handheld part, do not hold 
over the barrier when in use. 

Power on/off 

Push on the button      to turn on the instrument, and hold down for 
>1s to shut it off. 

Auto power off 

The instrument shuts off automatically after 20 minutes without 
buttons pushed. 

Deactivate Auto power off 

When the instrument is off, push and hold down       while turning 
the instrument on by pushing the     button. The text ”AoFF” is 
shown in the display for a few seconds before the instrument turns 
on. 

Auto ACA/DCA detection (Only Elma 3039) 

Briefly push the     button to switch between AC and 
DC current measurement. 

Push and hold the      button for >1s to return to 
“Auto” mode. 

Содержание 3037

Страница 1: ...Elma 3037 3039 Clamp Meter Elma 3039 https www elso sk product php id_product 3219 EAN 5706445900046 5706445900053...

Страница 2: ...ttery indicator Do not use Meter around explosive gas or vapor To reduce the risk of fire or electric shock do not expose this product to rain or moisture Individual protective equipment should be use...

Страница 3: ...shut it off Auto power off The instrument shuts off automatically after 20 minutes without buttons pushed Deactivate Auto power off When the instrument is off push and hold down while turning the inst...

Страница 4: ...shown in the display When the current exceeds 5A the instrument calculates the RMS value for 100ms Low pass filter When measuring AC push the button to enable Low pass filter LPF is shown in the displ...

Страница 5: ...rature 20 C 60 C 80 RH no batteries Temperature coefficient 0 2x accuracy C 18 C 28 C Pollution degree 2 Chock vibration MIL PRF 28800F for A class 2 instr Drop protection 1 2m drop to hardwood on con...

Страница 6: ...A 2 0 5 digits 1 5 5 digits 2 5 5 digits 300 0A 0 1A LCD displays 0 counts when the reading 20 counts Frequency response Only Elma 3039 50 400Hz Sine wave All specifications are valid from 5 to 100 of...

Страница 7: ...to repair this instrument The instrument contains no user serviceable parts The instrument should be turned in to Elma Instruments for service Cleaning Periodically wipe the case with a dry cloth and...

Страница 8: ...Elma 3037 3039 20 MD 20131023 v1 4...