• Inspection of high-frequency output power in all of the modes

These technical safety inspections should be entrusted only to properly trained personnel. The
manufacturer assumes no responsibility for improper changes or repairs carried out on the unit or its
accessories by unauthorized persons, and the warranty of the unit expires immediately.

III. CAUTION AND WARNING

3.0

CAUTION AND WARNING

3.0.1

CAUTION:

Do not use the Surgitron in the presence of flammable anesthetics or other
flammable gases, flammable liquids, or flammable objects.

3.0.2

CAUTION:

Electrical shock hazard. Do not remove cover. Refer to authorized personnel for
service.

3.0.3

CAUTION:

Surgitron may present a hazard to patients with pacemakers. Consult qualified
medical personnel.

3.0.4

WARNING: Hazardous electrical output. This equipment is for use only by qualified personnel.

3.0.5

WARNING: Never increase the power settings without first checking both the active electrode

and the neutral plate.

3.0.6

WARNING: The manufacturer-supplied accessories should be used to ensure the proper

operation of safety monitoring circuit.

To ensure safe and reliable operation, the Surgitron unit and the relevant accessories should be
checked for wear and tear before operation by qualified personnel.  

A maintenance contract with the manufacturer or one of its authorized service agents is
recommended after the warranty period.

3.1

SAFETY INSTRUCTIONS

Please review the below cautionary instructions to reduce the incidence of accidental burns.

3.1.1

The neutral electrode should be reliably attached with its entire area attached to the patient’s body
as close as possible to the operating field.

3.1.2

The patient should not come into contact with metal parts which are earthed or which have
appreciable capacitance to earth. The use of antistatic sheeting is recommended for this purpose.

3.1.3

Skin to skin contact (for example, between the arms and the body of the patient) should be
avoided.

3.1.4

When the Surgitron is used simultaneously with physiological monitoring equipment on the same
patient, the monitoring electrodes should be placed as far as possible from the surgical electrodes.
Needle monitoring electrodes are not recommended. Monitoring systems incorporating high-
frequency current-limiting devices are recommended.

3.1.5

CAUTION:  Interference caused by the Surgitron could adversely affect the operation of other

electronic equipment.

3.1.6

The cables to the surgical electrodes should be positioned in such a way that contact with the
patient or other leads is avoided. Temporarily unused active electrodes should be stored
segregated from the patient.

3.1.7

For surgical procedures where the HF current could flow through parts of the body having a

Содержание Surgitron 4.0 Dual RF/120 IEC

Страница 1: ...074 01 03...

Страница 2: ...1 Rear Panel 4 2 Front Panel V PREPARATION FOR USE Page 14 5 0 Preparation for Use 5 1 Operational Preparation and Functional Selection of the Surgitron Dual Frequency VI CLINICAL INFORMATION AND PRA...

Страница 3: ...zor sharp knife Electrocoagulation is a nonvolatilizing destruction of tissue cells by a radiofrequency wave The atraumatic nature of electrosection provides a noteworthy advantage The lack of trauma...

Страница 4: ...change without notice 2 1 GENERAL DESCRIPTION The Wolf ellman Surgitron 4 0 Dual RF 120 IEC enhanced capability Radiosurgical Generator described herein is a compact source of high frequency low temp...

Страница 5: ...Deeply modulated square wave with average to peak ratio approximately 35 Modulation occurring at a 60 or 50 Hz rate 2 4 4 FULGURATE Deeply modulated wave with average to peak ratio to 10 or less Modul...

Страница 6: ...WITH AIR OR WITH OXYGEN OR NITROUS OXIDE 2 6 6 The mode of operation is continuous operation with intermittent loading duty cycle 10 seconds on 30 seconds off 2 6 7 The permissible operational enviro...

Страница 7: ...aintenance contract with the manufacturer or one of its authorized service agents is recommended after the warranty period 3 1 SAFETY INSTRUCTIONS Please review the below cautionary instructions to re...

Страница 8: ...E The Wolf ellman Surgitron 4 0 Dual RF 120 IEC comprises some unique features which include retractable mode and power output level setting the waterproof dual footswitch for CUT and COAG modes and a...

Страница 9: ...er down push button for CUT COAG mode l During this power adjustment the handpiece acts as two button fingerswitch handpiece thus the power output intensity may be tested m Depress the SELECT touch sw...

Страница 10: ...should be set to a volume that can easily be heard by the user 4 1 5 Footswitch Receptacle This 9 pin receptacle accepts a Hypersonics Footswitch connector 4 1 6 Dual Footswitch Pedal The unit is ship...

Страница 11: ...pment For error codes information see 7 5 Error Codes Table 4 2 2 Fault LED Indicator It will be Red if any error is detected otherwise it should be Green 4 2 3 Neutral Plate Receptacle The neutral pl...

Страница 12: ...been selected See section 4 0 2 for Three Button Fingerswitch Handpiece application 4 2 12 Yellow Light Cut Coag Indicator The indicator illuminates when Cut Coag function has been selected See sectio...

Страница 13: ...Indicator for Coagulation Mode The default state is zero when On Off power switch is turned on See section 4 0 1 for special features of the power storage The output power will remain at the selected...

Страница 14: ...13 Figure 4 2 Numbers for Description of the Front Panel Control Elements...

Страница 15: ...to select the desired power level Depress the CUT Footswitch controller or Three Button Fingerswitch to activate the output power See section 4 0 2 for three button fingerswitch handpiece application...

Страница 16: ...ppreciate the tremendous advantages inherent in radiosurgery 6 1 1 Defining Good Technique Radiosurgery can create tissue damage if the heat is allowed to accumulate in the tissue to the point where e...

Страница 17: ...de is then brought into contact with the hemostat 1 or 2 inches from its tips When the Partially Rectified Modulated Waveform is turned on the walls of the vessel will be sealed making ligation unnece...

Страница 18: ...nt gentle contact with the tissue is performed until bleeding stops The same meat specimen used for cutting may be used for coagulating practice Effective coagulation is achieved when the treated area...

Страница 19: ...IONS 7 0 TECHNICAL SPECIFICATIONS 7 1 OUTPUT CHARACTERISTICS 7 2 SAFETY FEATURES All waveforms are interrupted at a 37 Hz rate to allow sampling with the power off 7 2 1 Standard Compliance IEC60601 1...

Страница 20: ...50 Hz 7 3 3 Power consumption at max output power 330 VA 7 3 4 Input current at max output power 1 30A 7 3 5 Rating of fuses in the main supply power T 2 0A 250V Two 7 4 SIZE AND WEIGHT 7 4 1 W x H x...

Страница 21: ...IGITAL SETTING 500 RESISTIVE LOAD 8 2 CUT MODE POWER OUTPUT VS RESISTIVE LOAD 8 3 CUT COAG MODE POWER OUTPUT VS RESISTIVE LOAD 8 4 HEMO MODE POWER OUTPUT VS RESISTIVE LOAD 8 5 FULGURATE MODE POWER OUT...

Страница 22: ...LAR MODES NOTE The above power chart is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used...

Страница 23: ...t is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection T...

Страница 24: ...art is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection...

Страница 25: ...t is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection T...

Страница 26: ...the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection The actual power as a function of load r...

Страница 27: ...the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection The actual power as a function of load re...

Страница 28: ...a com AUSTRIA Phone 01 405 62 19 Fax 01 405 51 51 45 Email info richard wolf at BELGIUM Phone 09 280 81 00 Fax 09 282 92 16 Email endoscopy richard wolf be FRANCE Phone 03 26 87 02 89 Fax 03 26 87 60...

Отзывы: