102 of 124 |
Mass Storage Class Tests
Ex
ami
n
er
U
se
r
G
u
ide
8.14
Case 10 Test
Test Summary
Examiner sends a series of mass storage class requests intended to confirm the case:
Ho <> Di: host wants to transfer data OUT, device wants to transfer
data IN.
Test Steps and Expected Results
1.
Examiner initializes the link to U0.
2.
Examiner issues several standard descriptor requests and places the device into the
configured state.
3.
Examiner issues an Inquiry command (Data Transfer Length = 36 bytes), a Test Unit
Ready command (Data Transfer Length = 0 bytes), and a Read Capacity command (Data
Transfer Length = 8 bytes).
Multiple Test Unit Ready commands may be issued until the device
reports it is ready.
4.
To induce the Case 10 behavior, Examiner issues a Read (10) with CBWDataTransferLength
=
BlockSize,
Transfer Length = 1 block, and Direction = Data OUT.
The test fails if any of the requirements listed below are not met:
The returned CSW must be 13 bytes in length.
The signature must be correct.
The Tag must match the Tag on the CBW.
The dDataResidue field must be equal to 0 Bytes.
The bStatus field must be equal to must be equal to 0x01 (Command Failed) or
0x02 (Phase Error).
Specification Reference
USB Mass Storage Device Compliance Specification TD1.14.
Содержание USB Explorer 280
Страница 1: ...USB Explorer 280 Examiner Compliance Test Suite User Manual Version 1 03 January 18 2013...
Страница 8: ...8 of 124 Examiner User Guide 10 Hub Tests 124...
Страница 24: ...24 of 124 Installing the Application Examiner User Guide The installation is complete...
Страница 124: ...124 of 124 Hub Tests Examiner User Guide 10 Hub Tests This section in process Contact Ellisys for more information...