PRE-SET THERAPY PROTOCOLS

Pre-set therapy protocols do not have variable adjustment parameters other than 
treatment amplitude (intensity). All functions are subject to treatment time.

Dual Therapy

 -

 Output base frequency for set dwell time, then abrupt jump to high 

frequency for set dwell time, then repeat. After treatment time reaches ½ total treat-
ment time, smooth ramp from 80 or 100Hz to 150Hz over set sweep time, then back.

Ramp

 -

 Smooth ramp from base frequency to high frequency over set sweep time, 

then back.

Abrupt 

-

 Output base frequency for set dwell time, then abrupt jump to high fre-

quency for set dwell time, then abrupt drop to base frequency, repeat.

Relax and Contract Stimulation (muscle stimulation)

 -

 Relax = On for 

set dwell time at pre-set base frequency and zero amplitude (intensity), then Con-
tract = fast amplitude ramp to amplitude (intensity) setting and maintain contraction 
for set dwell time at preset base frequency of 50Hz. Contraction indicated by
asterisk on screen. The amplitude cannot be increased during the contract state.

PRE-SET THERAPY PROTOCOLS

Protocol #

Function

Treatment 

Timer 

Base Freq. 

High Freq.

Dwell Time

Protocol 1

Abrupt 2/2

30 minutes

1 Hz

10 Hz

2 seconds

Protocol 2

Abrupt 6/6

30 minutes

1 Hz

20 Hz

6 seconds

Protocol 3

Abrupt 9/9

30 minutes

1 Hz

30 Hz

9 seconds

Protocol 1

Protocol 2

Protocol 3

10Hz

1Hz

2s

2s

20Hz

1Hz

6s

6s

30Hz

1Hz

9s

9s

Содержание Multi-Mode Stimulator

Страница 1: ...Carrier Frequency 4000 Hz Interference Frequency IF 4001 Hz to 4150 Hz Net Interferential Frequency Output Voltage 22 5 volts peak 0 45 volts peak to peak adjustable in 1 10 volt increments Output Cur...

Страница 2: ...device KEEP THIS DEVICE OUT OF THE REACH OF CHILDREN Follow your physician s or therapist s instructions as to which therapy presets to use electrode placement and suggested amplitude levels TABLE OF...

Страница 3: ...y that does change depending on the therapy protocol When these two stimulation frequencies cross and interfere with each other within the body the two frequencies subtract resulting in a third freque...

Страница 4: ...an or therapist The amplitude or intensity of the treatment is patient controllable INTRODUCTION TO THE MULTI MODE STIMULATOR The Multi Mode Stimulator is a high quality advanced technology medical de...

Страница 5: ...ns etc 7 Stimulation should not be applied over or in proximity to cancerous lesions 8 Stimulation should not be used whenever pain syndromes are undiagnosed until etiology is established PRECAUTIONS...

Страница 6: ...Electronic monitoring equipment such as ECG monitors and alarms may not operate properly when interferential stimulation is in use in the management of pain patients 14 The Multi Mode Stimulator shou...

Страница 7: ...tude Lockout To prevent an accidental increase or decrease of the treat ment amplitude intensity the Multi Mode Stimulator has an intra therapy amplitude lock feature Sixty seconds after the last incr...

Страница 8: ...e quency for set dwell time then abrupt drop to base frequency repeat Relax and Contract Stimulation muscle stimulation Relax On for set dwell time at pre set base frequency and zero amplitude intensi...

Страница 9: ...sec high 30 minutes 80 Hz 150 Hz 12 seconds Protocol Function Treatment Timer Base Freq High Freq Sweep Time Protocol 6 Full Sweep 6 seconds 30 minutes 1 Hz 150 Hz 6 seconds Protocol 7 Full Sweep 15 s...

Страница 10: ...9 Dual MED Abrupt shift Ramped shift 15 minutes 15 minutes 1 Hz 80 Hz 10 Hz 150 Hz 6 seconds 6 seconds Protocol 10 Dual HIGH Abrupt shift Ramped shift 15 minutes 15 minutes 10 Hz 100 Hz 20 Hz 150 Hz 6...

Страница 11: ...col 13 Muscle Simulation 20 minutes 50 Hz 50 Hz 6 sec ON 6 sec OFF Protocol 14 Muscle Simulation 20 minutes 50 Hz 50 Hz 5 sec ON 10 sec OFF Protocol 11 Protocol 12 Protocol 13 Protocol 14 3s 3s Contra...

Страница 12: ...to the elec trodes and connect the opposite end to the therapy cable connector on the top of the stimulator 3 Turn the stimulator ON In a few moments you will see a screen asking you to make a therapy...

Страница 13: ...your healthcare professional THERAPY STATUS SCREEN TYPICAL DISPLAY SCREENS SELECT PROTOCOL AMP LOCK PRESS ENTER TO UNLOCK THERAPY COMPLETE A090 P12 TR 07 Amplitude Setting A setting of 90 equals 9 vol...

Страница 14: ...replacement lead wires Simulation Felt but the Device is Off Device is defective DO NOT USE please contact manufacturer for repair or replacement TYPICAL DISPLAY SCREENS Continued SYSTEM FAILURE CONTA...

Страница 15: ...se Remove the batteries from the device during storage WARRANTY INFORMATION Elite Medical Supply of NY warrants the Multi Mode Stimulator against any defects in materials and workmanship for a period...

Страница 16: ...Manufactured by Elite Medical Supply of NY LLC 1900 Ridge Road Suite 125 West Seneca NY 14224 866 712 0881 For questions regarding proper use or to request more supplies please contact Elite Medical S...

Отзывы: