PRE-SET THERAPY PROTOCOLS
Pre-set therapy protocols do not have variable adjustment parameters other than
treatment amplitude (intensity). All functions are subject to treatment time.
Dual Therapy
-
Output base frequency for set dwell time, then abrupt jump to high
frequency for set dwell time, then repeat. After treatment time reaches ½ total treat-
ment time, smooth ramp from 80 or 100Hz to 150Hz over set sweep time, then back.
Ramp
-
Smooth ramp from base frequency to high frequency over set sweep time,
then back.
Abrupt
-
Output base frequency for set dwell time, then abrupt jump to high fre-
quency for set dwell time, then abrupt drop to base frequency, repeat.
Relax and Contract Stimulation (muscle stimulation)
-
Relax = On for
set dwell time at pre-set base frequency and zero amplitude (intensity), then Con-
tract = fast amplitude ramp to amplitude (intensity) setting and maintain contraction
for set dwell time at preset base frequency of 50Hz. Contraction indicated by
asterisk on screen. The amplitude cannot be increased during the contract state.
PRE-SET THERAPY PROTOCOLS
Protocol #
Function
Treatment
Timer
Base Freq.
High Freq.
Dwell Time
Protocol 1
Abrupt 2/2
30 minutes
1 Hz
10 Hz
2 seconds
Protocol 2
Abrupt 6/6
30 minutes
1 Hz
20 Hz
6 seconds
Protocol 3
Abrupt 9/9
30 minutes
1 Hz
30 Hz
9 seconds
Protocol 1
Protocol 2
Protocol 3
10Hz
1Hz
2s
2s
20Hz
1Hz
6s
6s
30Hz
1Hz
9s
9s