- 17 -
7.3
Default Parameter / Calibration
7.3.1 Calibration
(The Sensor must be installed on the Magnetic tape!)
Ö
Switch off the unit.
F
press this button
Ö
While pressing this button switch the unit on again
Here, the calibration of the sensor is triggered and "CAL 0" is shown in the display. Now, move the sensor
slowly in one direction on the tape. After the calibration (display "CAL 0"… "CAL 4") is the display again in
normal mode.
7.2.2 Loading the Default Parameter
(The Sensor must be installed on the Magnetic tape!)
Ö
Switch off the unit.
Incr/abs
press this button
Ö
While pressing this button switch the unit on again
Here, all parameters will be reset to default parameters. Also herewith the calibration of the sensor is
triggered and "CAL 0" is shown in the display. Now, move the sensor slowly in one direction on the tape.
After the calibration (display "CAL 0"… "CAL 4") is the display again in normal mode.
Structure and Function
Содержание Z16 Series
Страница 1: ......