SW 5250A•SW 3500A•SW 1750A
Operation
Operation Manual
3-5
3.3.2 PROGRAM
Menu
The
PROGRAM
menu allows you to:
•
Select a phase
•
Program amplitude
•
Program current limit
•
Program frequency
•
Program phase offset (phase offset represents a phase lead with respect to the
internal reference)
•
Select a function to be output on the selected phase.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
PHASE A (or B or C)
AMPL
V
CURL
A
FREQ
Hz
ANG
Deg
FUNC
MAIN MENU
OUTPUT [OFF]
θ
PROGRAM Menu Status and Information Data:
Menu Item
Status and Information Line
Range/Default
PHASE A
[<enter> for next phase]
Phases A, B, C
AMPL
[program output RMS/DC voltage] *
0–156 V or 0–312 V
(default: 0.00V)
CURL
[program RMS/DC current limit]
0–13 A or 0–6.5 A (default: 5.00A
dependent on voltage range)
FREQ
[program output frequency]
**
40–5000 Hz (default: 60 Hz)
θ
ANG
[program phase angle offset] ***
0–360° (default: 0°, 120°, 240°
dependent on phase selected)
FUNC
[program output waveform]
****
Default: Sine
Содержание SmartWave SW1750A
Страница 2: ... ...
Страница 4: ...Committed to Quality Striving for Excellence ii ...
Страница 5: ...This page intentionally left blank iii ...
Страница 8: ......
Страница 31: ...SW 5250A SW 3500A SW 1750A Installation Operation Manual 2 5 Figure 2 3 SmartWave Rear Panel ...
Страница 66: ...Operation SW 5250A SW 3500A SW 1750A 3 22 Operation Manual Table 3 1 Sequence Execute Menu ...
Страница 90: ......
Страница 94: ...Standard Waveforms SW 5250A SW 3500A SW 1750A 4 4 Operation Manual This page intentionally left blank ...