![ELGA PAF0076-01 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/elga/paf0076-01/paf0076-01_configuration-and-installation-manual_2388421020.webp)
20
4 x PURELAB Chorus 3 (30lph) + 60 or 100 liter Storage
Reservoir
Hydraulic / Water Connections
Electrical / Communications Connections
Communication / Power Connection Details
Colour
Description
System 1
Connection No.
System 2
Connection No.
System 3
Connection No.
System 4
Connection No.
Storage
Reservoir
Black
Power Cable
6
-
6
-
Red
Internal Comm’n Cable
J2
J2
-
-
-
Red
Internal Comm’n Cable
-
-
J2
J2
-
Grey
Internal Power Cable
TH4
TH4
-
-
-
Grey
Internal Power Cable
-
-
TH4
TH4
-
Black
External Comm’n &
Power Cable
7
-
7
-
-
Black
External Comm’s &
Power Cable
-
-
-
7
J2
Parts / Installation Kits Required
Part No
.
Description
QTY
Supplied with
System
Ordered
Separately
CEPWGU303754
Power Supply
2
-
✓
LA762
Standard Installation Kit
4
✓
-
LA773
Reservoir Installation Kit
1
✓
-
LA777
Integrated Installation Kit
2
-
✓
LA778
Side By Side Installation Kit
1
-
✓
LA652
Pressure Regulator
With Boost Pump – Fitted when inlet pressure >2.0bar (30psi).
Without Boost pump – Fitted when inlet pressure >6.0bar (90psi).
2
-
✓
System 2
7
Power
Supply
.
6
J2
System 1
System 4
System 3
7
7
.
6
1
4
1
1
4
Drain
3
3
5
4
4
1
1
3
3
Inlet
Outlet
Inlet
Outlet
J2
J2
TH4
TH4
J2
J2
TH4
TH4
Storage
Reservoir
External
Communication
& Power Cable
External
Communication
& Power Cable
4a
Outlet