8. SANITIZATION PROCEDURES
8.1 Liquid sanitization
The normal sanitization procedure for the
MEDICA
unit is to sanitize
the Reverse Osmosis (RO) module and associated pipework (
8.1.1
). If
required, the unit and docking vessel/reservoir may also be sanitized
(
8.1.2
).
8.1.1 RO sanitization procedure
The unit is sanitized to reduce the growth of microbiological
contamination within the unit. The
MEDICA-R 7/15 US
has a built-in
sanitization port, which allows the sterilant to be introduced into the
water feeding the RO. Please read this entire section to become
familiar with the procedure before you start.
The sanitization frequency required is dependent on the feedwater,
local environment, usage patterns and application. As a general rule,
ELGA LabWater recommends sanitization of the RO at 6 monthly
intervals. However, the period between sanitizations could be
extended in particular circumstances. For example, microbial growth
will usually be lower with clean feed water, well-maintained pre-
treatment, low temperatures and heavy usage.
WARNING! ALWAYS WEAR RUBBER GLOVES, APRON AND
FACEMASK. DO NOT BREATHE FUMES OR ALLOW
TO COME IN CONTACT WITH SKIN OR EYES.
ALWAYS FOLLOW RECOMMENDATIONS FOUND IN
THE MANUFACTURERS MATERIAL SAFETY DATA
SHEET AND ANY APPLICABLE OSHA STANDARDS
FOR THE CHEMICAL BEING USED.
WARNING! LABEL
THE
MACHINE
WITH
APPROPRIATE
WARNING SIGNS SUCH A
S “DO NOT USE/
CONTAINS STERILANT” (NOT PROVIDED)
Minncare Cold Sterilant is a Peracetic and Hydrogen Peroxide based
solution.
Refer to the Minncare label for additional information and follow all
applicable directions for use on the manufacturer’s label in conjunction
with the following instructions.
Step 1 - Start Sanitization Cycle
1. ENSURE that the reservoir level indication on the
graphics display is showing >40%. If display
shows >70% or 100%, dispense water until
display changes to >40%.
2. PRESS the PROCESS button to stop the
process.
3. TURN OFF the electrical power supply.
4.
APPLY a suitable warning label such as “DO
NOT USE / CONTAINS STERILANT”.
5. RELIEVE residual pressure by opening the
dispense tap and directing water to drain.
6. CLOSE the dispense tap.
7. DISCONNECT outlet tube at reservoir inlet and
re-direct to a large container (>5litres)
– refer to
Process flow diagram on p6.
8
0.05
Process On