MEDICA-R 7/15 US
Operator Manual
ELGA
Page 26
MEDICA-R 7/15 US
Version 2
– 10/15
7.1 Replacing the LC140 pre-treatment cartridge
The pre-treatment cartridge should be replaced when indicated by the
change reminder.
Step 1 - Switch unit off
1. SWITCH the
MEDICA-R 7/15 US
off at the power
switch at the top left hand side of the unit.
2. ENSURE pressure has dissipated from the unit
by waiting several minutes before proceeding.
Step 2 - Remove pre-treatment cartridge
1. OPEN front door.
2. IDENTIFY the pre-treatment cartridge (LC140).
3. REMOVE the tube at the elbow at the bottom of
the cartridge, by pushing back the retaining collet
on the push fit connector and withdrawing the
reducer.
4. REMOVE the tube from the elbow at the top of
the cartridge.
5. REMOVE exhausted cartridge from retaining
clips and discard.
Note: The consumable is non-hazardous. Dispose of as
ordinary waste, observing all local and national
regulations.
Step 3 - Replace the pre-treatment cartridge
1. UNPACK new cartridge and REMOVE the two
protective transit plugs sealing the inlet and outlet
connection.
2. SECURE the new cartridge into its retaining clips
ensuring the cartridge is the correct way up.
3. REFIT the inlet tubing into the bottom of the
cartridge by pushing the tube into the elbow
connector until locked and held by the retaining
collet.
4. REFIT the outlet tubing to the top of the cartridge.
5. RESET pre-treatment cartridge timer
. (See
Section
5.5
-
Setting
up
Display
and
Replacement Timer
– Step 3).
6. PRESS the PROCESS button to start the unit.
7. CHECK the unit for leaks and CLOSE front door.