background image

SPOT V2I6 Manual.docx 

12/01/2018 

 

 

Spot Series Lasers 

(Short Pulse Optical Technology) 

 

 

 

Elforlight Ltd 

Nene House 

Drayton Way 

Daventry 

Northants 

NN11 8EA 

UK 

 

Tel +44 1327 300069: Fax +44 1455 552974 

 

www.elforlight.com 

 

Содержание Spot Series

Страница 1: ...V2I6 Manual docx 1 12 01 2018 Spot Series Lasers Short Pulse Optical Technology Elforlight Ltd Nene House Drayton Way Daventry Northants NN11 8EA UK Tel 44 1327 300069 Fax 44 1455 552974 www elforligh...

Страница 2: ...ument or the product which it accompanies The contents of this manual are subject to change without notice Please contact the factory to ensure you have the latest revision of this document before emb...

Страница 3: ...1 2 Laser safety 1 3 Recommended safety precautions during laser operation 1 4 System handling mounting and heat sinking 1 5 Basic Operation Connections Start up procedure 1 6 Operation with power var...

Страница 4: ...ll SPOT lasers emit laser radiation which is potentially hazardous and it is therefore strongly recommended that before operating the laser for the first time users should thoroughly familiarise with...

Страница 5: ...esponsibility to ensure full compliance with these standards Emission The laser emits CW or pulsed radiation at 1064nm 532nm and 355nm or as otherwise specified Powers up to 2Watts may be generated Th...

Страница 6: ...ser when not in use and or unattended and stored in a safe place Eye and skin exposure to direct or scattered laser radiation is hazardous and should be considered potentially extremely harmful Suitab...

Страница 7: ...on and disconnection of the system sub units Please contact Elforlight for recommended anti static procedures Connect the laser head to the power supply unit using the 37 way D connector socket on the...

Страница 8: ...n The keyswitch may be switched ON at any time during this 4 minute period without prolonging the next stage of start up After the above 4 minute warm up period upon turning the key switch or immediat...

Страница 9: ...ould be configured with the parameters listed below 9600 baud 1 start bit 8 data bits no parity and 1 stop bit No hardware handshaking No flow control To operate the laser using power variation by RS...

Страница 10: ...h voltage switch in the HV module A further adjustable DC supply to drive the high voltage supply in the HV module A cooling fan Interconnecting wiring to bring limited functionality of the 18 04 supp...

Страница 11: ...sequence for details 3 HV adjust potentiometer if fitted Adjusts the high voltage applied to the Pockels Cell to allow operation over a wide range of repetition rates See test data for settings 4 TTL...

Страница 12: ...way Neutrik connector Contact closure between pins 1 and 3 is required to enable laser operation Laser is supplied with a connector with a shorting link in place 4 Laser head 37 way D connector Provi...

Страница 13: ...er Increment Iset parameter by one Decrement Iset parameter by one Request status Q Request software version number RS232 protocol 9600 baud 1 start bit 8 data bits no parity 1 stop bit No hardware ha...