![elexxion claros pico Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/elexxion/claros-pico/claros-pico_user-manual_2388344016.webp)
claros pico user manual V 2.1
page 16 of 34
6
Operation
6.1
Commissioning
Please wait at least 30 minutes after moving from a cold environment to a warm environment before switching the
device on to prevent condensation forming on electronic components.
1.
Please ensure that the device is placed in a secure and stable position on the stand supplied. The max. load of
the stand is 500 g. Do not use the pico without the stand.
2.
claros pico is fitted with rechargeable batteries. The batteries must be charged before the device is first brought into
service. A full charging cycle takes up to 3.5 hours. In order to charge the device, connect the mains adapter plug
supplied to the socket on the lower front side of the claros pico device.
3
The charging level is displayed in the top right-hand corner of the display. The indicator will flash during charging.
The battery charging level is permanently displayed after charging is complete.
4
A visual warning is displayed when only a limited amount of battery charge is left. The charging level indicator will
flash rapidly. Active operation of the laser is possible for a maximum of 5 minutes after this time.
5
The operating time of the device when fully charged depends on the power output used. When all programs are
used evenly, it is possible to operate the device for around 45 minutes.
6
The foot switch is brought into service by plugging the foot switch connection lead into the socket provided on the
left-hand side of the claros pico casing. Push the foot switch plug in until it reaches the stop of the device socket.
Take care, that the red dots are facing each other.
Attention! Activating the foot switch while the unit is activated emits class 4 laser radiation.