14
80A2E22
3
For TIVG-V15:
To re-adjust, remove the cover and the adjusting dial is then
visible at the top. Put a screwdriver in the slot of the adjusting dial and gently
turn the dial to the desired position. Your will find that you by doing this are
moving the plate, where the micro switch is mounted mechanically.
Adjusting of high/low flow alarm for TIVG-V15
Adjusting dial
Figure 8
If you have the possibility to check against a flow meter in the system, you
will get the best on site adjustment of the switch point.
Reinstall the cover at the top with the two screws and start up the process
again.
3.4
Change of Flow Direction
Empty the pipe system so that it is un-pressurized and has no flow!
At the time of ordering, you must specify in which direction the Flow Monitor
shall be mounted i.e. from which side is the flow entering the Pipe Section and
how you would like to read the scale. (Please refer to fig. 6 below for alternatives.)
If, for some reason, the Flow Monitor is ordered with the wrong flow direction,
it is possible to change this in the field.
3.4.1
Flow direction Selector
The threaded brass version DN40 and the flanged stainless steel version are
delivered with a flow direction selector that can be used for both directions.
To change the direction, loosen the four (4) hexagon screw, which hold the di-
aphragm housing to the Pipe Section.
1,5
17
4A
2E
Содержание TIVG-100F
Страница 1: ...Manual Eletta Flow Monitor TIVG S Modell S2 o05 2021 04 20 80A2E22...
Страница 19: ...19 80A2E22 5 1 Measuring Ranges Tables 5 MC Measuring constant on TIVG S25...
Страница 20: ...20 80A2E22 5 5 2Weight and Dimensions TIVG 15 and 25R TIVG V15 40R...
Страница 21: ...21 80A2E22 5 TIVG S25 15 and 25R TIVG S25 40R...
Страница 22: ...22 80A2E22 5 TIVG S25F...
Страница 23: ...23 80A2E22 6 6 1 TIVG V15 15 and 25R...
Страница 24: ...24 80A2E22 6 6 2 TIVG V15 40R...
Страница 25: ...25 80A2E22 6 6 3 TIVG S25 15 and 25R...
Страница 26: ...26 80A2E22 6 6 4 TIVG S25 40R...
Страница 27: ...27 80A2E22 6 6 5 TIVG S25F...