Adjustment of switch point 

All the Indicating units on Eletta  Flow  Monitors are tested and calibrated 

according to the customers’ orders before shipping. If the customer does not 

specify a desired switch point for the flow alarm, the TIVG-S25  Monitor’s 

micro switches are preset to trip at the min-  and max flow value.  For the 

TIVG-V15 the 

micro switch is preset to trip at 50% (decreasing flow) of the 

ordered max flow value, for example a 10 - 20 l/min flow range would have the 

switch set to trip at 15 l/min, decreasing flow.

 

Please note!! 

We have calibrated each Indicating unit of the Flow Monitor in 

our flow rig and set the switches according to the Flow values we achieve in 

the rig under good conditions. We  must stress that under actual  field 

conditions,  the  flow  profile  can  be  different  from  the  one  in  our  flow  rig 

depending on valves, hoses, bends or other obstructions and therefore the 

switching can be off from our preset values. There is a possibility to adjust the 

switch/alarm points in the field by adjusting the micro switches’ position me- 

chanically.  
 

For TIVG-S25

: To readjust, remove the two screws that hold the cover at the 

top of the blue housing. The two adjusting dials are then visible through the 

opening. Underneath the removed cover, you will find a small tool necessary 

to use in order to change the adjusting dials position. 

The adjusting dials are marked the same as the scale in the front and this mar- 

king can be used to approximately find the right switch/alarm point for the ac- 

tual application. Put the tool inside the drilled hole on the top of the adjusting 

dial and gently move the dial sideways to the desired position. If the two ad- 

justing dials are set to the same position, the micro switches will trip at the 

same time. If possible, use the left adjusting dial for the high flow alarm (hig- 

her end of the scale) and the right for the min. flow alarm (the lower end of the 

scale), in order to get the best accuracy. If you use the alarms the other way 

around, the spring mechanism inside the diaphragm housing will be affected 

with lower accuracy as a result, so please try to avoid that. Repeat the 

procedure for the next adjusting dial and then put the tool back in its bracket at 

the cover.  

 
 
 
 
 

Figure 7 

Содержание TIVG-100F

Страница 1: ...Manual Eletta Flow Monitor TIVG S Modell S2 o05 2021 04 20 80A2E22...

Страница 2: ...docu mentation such as manuals drawings leaflets etc are not permitted without a written permission from Eletta Flow AB Sweden These data are only disclosed to you under permission of limited use with...

Страница 3: ...3 4 Change of Flow Direction 14 3 4 1 Flow Direction Selector 14 3 5 Change of Dial Orientation 15 Trouble shooting 16 4 1 Verification of Flow 16 4 2 Electrical connections 17 4 3 Spares 18 Tables 1...

Страница 4: ...s mainly i e the Pipe Section and the Control Unit The Pipe Section is the part that is to be mounted in the process pipe and the Control Unit is mounted directly standard or remote on to the Pipe Sec...

Страница 5: ...alarm set points are within the chosen flow range Repeatability 2 actual Pressure Max 16 bar PN16 DN250 400 10 bar PN10 Min Line pressure of apx 0 7 1 0 bar 10 14 5 PSI is required for proper operatio...

Страница 6: ...nless steel units Fluorinated rubber FPM and for brass unit DN40 Nitril HNBR Spacer G40 only The spacer that holds the orifice plate inside the pipes section and they are made of Polyamide plastic PA...

Страница 7: ...oint in the piping system if you are measuring liquids and at the highest point if you are measuring gases Also check if the planned flow direction in the system matches the one indicated on the Monit...

Страница 8: ...e of the fact that it is practically impossible to predict when the flow is stable after disturbances in the piping so this must serve as a guideline only The straight runs must be free from valves be...

Страница 9: ...les from the micro switch electric terminal On the Brass G40 Pipe Section loosen the four 4 hexagon screws that hold the diaphragm housing do not remove the blue housing at any time to the Pipe Sectio...

Страница 10: ...ase note that you can have two alternative moun tings of the cable gland depending on what side you want to enter with the ca bles The terminal block connections are described in fig 5 below A groundi...

Страница 11: ...ally via a rubber diaphragm and linked to an outside of the process liquid gas mechanism This mechanism transforms the movement into a Flow rate value shown on the dial All Indicating units of the Ele...

Страница 12: ...achieve a new flow range Please follow the above instructions for the brass model for dismounting the whole orifice plate with holder Remove the Control Unit from the old Pipe Section orifice section...

Страница 13: ...d by adjusting the micro switches position me chanically For TIVG S25 To readjust remove the two screws that hold the cover at the top of the blue housing The two adjusting dials are then visible thro...

Страница 14: ...p the process again 3 4 Change of Flow Direction Empty the pipe system so that it is un pressurized and has no flow At the time of ordering you must specify in which direction the Flow Monitor shall b...

Страница 15: ...ain mounting position in the piping system the large visible dial can be mounted in several positions to suit the application please refer to fig 7 If you need to change the orienta tion in the field...

Страница 16: ...3 4 4 1 Verification of flow We would like to stress the fact that all Indicating units on the Eletta Flow Monitors are calibrated and adjusted individually in water in a specially purpose built cali...

Страница 17: ...If so move the Monitor further away to achieve enough straight runs You can of course verify the flow in the Eletta Flow Monitor versus an other flow meter in the system or take the Monitor out and p...

Страница 18: ...r flow range and liquid and order a new Unit from us We will ship you the complete Control Unit with diaphragm housing and you can then easily fit the new Control Unit to your existing Pipe Section wi...

Страница 19: ...19 80A2E22 5 1 Measuring Ranges Tables 5 MC Measuring constant on TIVG S25...

Страница 20: ...20 80A2E22 5 5 2Weight and Dimensions TIVG 15 and 25R TIVG V15 40R...

Страница 21: ...21 80A2E22 5 TIVG S25 15 and 25R TIVG S25 40R...

Страница 22: ...22 80A2E22 5 TIVG S25F...

Страница 23: ...23 80A2E22 6 6 1 TIVG V15 15 and 25R...

Страница 24: ...24 80A2E22 6 6 2 TIVG V15 40R...

Страница 25: ...25 80A2E22 6 6 3 TIVG S25 15 and 25R...

Страница 26: ...26 80A2E22 6 6 4 TIVG S25 40R...

Страница 27: ...27 80A2E22 6 6 5 TIVG S25F...

Страница 28: ...be mixed with general household waste For proper recycling please take these products to a designated collection point where they will be accepted free of charge Please contact your local authority or...

Отзывы: