Page 27 of 40 M8144
Issue 2.0
7.8. Printer setu-up.
PR2000S. Model printer is set as follows:.
The PR2000S serial printer is set up for compatibility with the IA. The Baud rate on
the PR2000S printer is factory set at 9600.
If using a printer that has been used from a previous model of IA the baud rate must
be changed to 9600.
To make local settings on the printer the printer must be put in ‘program mode’.
Press the ‘
’ and the ‘1’ button simultaneously. The printer will now enter
‘PROGRAMMER MODE’.
The following text will be printed.
“DATA/PAR’Y:- 8/NONE”
Press the ‘PAPER’ feed button.
The following text will be printed.
“Xon/Xoff:- ON”
Press the ‘PAPER’ feed button.
The following text will be printed.
“BAUD RATE:- 9600”
Press the ‘
’ and ‘PAPER’ feed buttons simultaneously.
The PR2000S will produce a print of the stored settings which will be as follows:
SETUP COMPLETE
:-
Data/Par’y
:-
8/NONE
Xon/Xoff
:-
On
Baud rate
:-
9600
Country
:-
UK
Print mode
:-
Text
Auto-off
:-
5 min
Emulation
:-
Standard
DTR
:-
Normal
Interface
:-
Serial
Mechanism
:-
M190