Page 26 of 40 M8144
Issue 2.0
7.7.15. The display will now ask for the certified chemical melt value to be entered.
. Enter the chemical reference value using the numeric
keypad. Press the return key to accept the entered value.
7.7.16. A preprogrammed algorithem will now calculate the PID curve for all melt values.
When that has been done the display will say
.
7.7.17. Remove the calibration key and powerdown the IA. It may now be powered back up
again ready for use. Calibration settings will now take effect.