A993
OTS5000
23
7. To change the thickness setting, press the START/STOP button once to stop movement if in
motion. Change the setting using the appropriate keypad buttons, then press the
START/STOP button again. The blade arm will return to its reverse limit, step down the new
thickness amount, and begin slicing.
8. Pressing the HOME button at any time will return the blade arm to its furthest up and back
position.
Foot Pedal Operation
1. Turn the power switch off if applicable
2. Ensure that the foot pedal is installed correctly per section 2.2.
3. Activate power using switch on the back panel. (Indicated by the l side of the rocker switch
pressed in.)
4. Press the START/STOP button to activate multiple slice mode. (The foot pedal needs to be in
this mode to function correctly.)
5. The foot pedal is used in place of the FORWARD keypad button in multiple slice mode above.
6. Pressing the foot pedal down begins blade arm advance and oscillation. Lifting off stops all
movement.
2.5
SCHEDULED MAINTENANCE
The surfaces of the EMS 5000 can be cleaned of any salt buildup etc. by washing with soap and
water. DO NOT use phenol or ketones to clean.
A yearly performance of the Functional Checkout in section 2.3 should be performed to ensure
function and calibration. If the unit fails any part of this functional test, contact the EMS Repair
Department at 215-412-8400