SmartVest
SQL Airway Clearance System
www.smartvest.com
17
3. To adjust the following settings:
•
Frequency
- Press or to the left or right of the frequency display.
Adjust in 1 Hz increments from 5 to 20 Hz.
•
Pressure
- Press to the left or right of the pressure display. Adjust
in
increments of 5% from 10%-100%.
•
Time
- Press or to the left or right of the time display. Adjust in 30
second increments from :30 - 60:00 minutes.
NOTE:
Frequency, pressure, and time cannot be changed during a Program
session.
4. Press to save the settings for this interval and move to the next column.
Note:
The currently displayed columns are underlined in the top-right corner of
the screen. Only three of the 6 (A-F) programmable columns can be shown at a
time. The blue column is being adjusted. The column to the right is next.
5. Repeat this process for columns B-F or for as many intervals, as needed.
NOTE:
Any columns left as 0:00 will be skipped during treatment.
6. Press to save the displayed values as the new default values (see
Starting the Treatment Protocol).
Repeat this same process to set additional Program modes, as needed.
SQL Air Pulse Generator
Displayed
columns
Next
editable
column
Currently
being
adjusted