www.smartvest.com
SmartVest
SQL Airway Clearance System
16
SQL Air Pulse Generator
2. To adjust the following settings:
•
Frequency
- Press or to the left or right of the frequency display. Adjust in
1 Hz increments from 5 to 20 Hz.
•
Pressure
- Press to the left or right of the pressure display. Adjust
in
increments of 5% from 10%-100%.
•
Time
- Press or to the left or right of the time display. Adjust in 30 second
increments from :30 - 60:00 minutes.
NOTE:
Frequency and pressure can be changed at any time during a Manual session.
Time can only be changed when the unit is not pulsing.
4. Press to save the displayed values as the new default values (see Starting the
Treatment Protocol).
Setting Program Mode
Steps:
1. From the Home screen, press Program-1, Program-2, or Program-3 to
select that mode.
2. From the selected Program mode (PGM-1, PGM-2, PGM-3) screen, press to
select the first column of the treatment protocol.