4.Lift the probe from the surface and adjust the
displayed reading down
(-)
or up
(+)
to the known
thickness value of the thin item. Press to accept
this value.
5.Repeat steps 2 - 4 for the thicker item.
When selected, the icon will appear and the
current calibration settings are “locked” to prevent
further user adjustments.
User Mode Calibration
There are applications where 1- or 2-point
calibrations will not work satisfactorily such as when
measuring magnetic coatings, or conductive
coatings over non-metals. For these applications the
included PosiSoft software can download special Cal
settings.
Calibration Memory
Calibration Memory
It is often convenient to retain a particular
calibration adjustment before making another. Then,
if you return to that part, the corresponding
calibration
setting
can be restored.
A “setting” is any calibration adjustment. The
PosiTector
6000
always displays the currentcalibra-
tionsetting(ex.
Cal 3
) in the upper right corner of the
display.
The setting called
Cal 1
has unique features. It can be
adjusted but never deleted, and is always made
active with factory settings after a
Reset
(see pg.15).
-creates a new calibration setting using the next
available number (Maximum of 10).
By default, these
new Cal settings are initially created with the Gage’s
factory settings.
This is indicated with the icon
9