Attenuation Ratio . . . . . . . . . . . . . 100x (1:100)
Bandwidth . . . . . . . . . . . . . . . . . . see table
Rise Time . . . . . . . . . . . . . . . . . . . see table
Input Resistance . . . . . . . . . . . . . . see table (used with oscilloscope with 1 MΩ input)
Input Capacitance . . . . . . . . . . . . . see table
Compensation Range . . . . . . . . . . see table
Max. Input Voltage . . . . . . . . . . . . 4000 V CAT I (DC + peak AC)
Derated with Frequency (see Derating Curve)
Pollution Degree 2
Max. Operating Temp . . . . . . . . . . 0° to +50° C
Humidity . . . . . . . . . . . . . . . . . . . . 85% RH or less (at 35° C)
Cable Length . . . . . . . . . . . . . . . . see table
Definitions:
Measurement Category I (CAT I) is for measurements performed on circuits not directly connected to mains.
Measurement Category II (CAT II) is for measurements performed on circuits directly connected to the low voltage in-
stallation. Examples are on household appliances, portable tools and similar equipment.
Pollution Degree 2 refers to an operation environment where normally only dry non-conductive pollution occurs
Temporary conductivity caused by condensation must be expected.
Low Voltage Directive
(LVD)
2014/35/EU
EN 61010-031:2015
Type C
Specifications [EC Declaration of Conformity]
0 V
1000 V
2000 V
3000 V
4000 V
5000 V
DC
100 MHz
1 MHz
10 MHz
Voltage vs Frequency Derating Curve
WARNING
.
Do not apply to the input any potential that exceeds the maximum
ratings of the probe .