3.12.3 Automatic compensation for cable resistance
This compensation is obtained by holding down the HOLD key when the continuity test or measurement resistance
function is selected.
When the key is released, when the display shows zero, the correction value is memorised in the clamp.
If the value measured is higher than 2
W
)
this correction is stopped and the memorised correction value is reset to
zero.
This correction is prohibited in MIN/MAX mode.
3.12.4 Automatic compensation of current measurement zero
This compensation is obtained by holding down the HOLD key when the current measurement function is selected.
When the key is released, when the display shows zero, the correction value is memorised in the clamp.
If the value measured is higher than 6 A, this correction is stopped and the memorised correction value is reset to zero.
This correction is prohibited in MIN/MAX mode.
3.12.5 Manual selection of AC, DC or AC+DC mode
By default the clamp switches to AC or DC mode automatically (AC or DC symbol flashing) for the V and A functions.
Short successive pressures on the yellow key are used to select manually AC measurement, DC measurement of compound
signals (AC+DC) and return to automatic mode. When the mode is selected manually, the AC, DC or AC+DC symbol is
fixed. The selection of this manual mode is impossible in MIN/MAX or HOLD mode.
3.12.6 3.12.6 Selections possible in continuity function
By default the clamp is in continuity function (
) for the position corresponding to the rotary switch. Successive presses
on the yellow key enable selection of the resistance measurement (
W
), then the semi-conductor test function ( )
and return to the continuity function (
)
3.12.7 Selection OF INRUSH function
This is done in function A (AC) by pressing first on the MIN/MAX key then on the yellow key.
Consultation of the values corresponding to this function is possible by pressing first on the HOLD key then by short
successive presses on the yellow key (see § 3.7).
Quitting this function is possible by short presses on the MIN/MAX key.
3.12.8 Suppression of automatic stop
Hold the HOLD key pressed and bring the rotary switch from the OFF position to the position The unit emits a double
beep then the symbol flashes. The selected configuration is memorised when the key is released: the
symbol remains lit continuously. The unit returns to automatic stop mode when the switch goes to the OFF position
.
Lapparecchio emette un duplice bip quindi il simbolo lampeggia.
La configurazione scelta viene memorizzata quando si rilascia il tasto : il simbolo è acceso fisso.
Lapparecchio ritorna in modo arresto automatico quando il commutatore passa sulla posizione OFF.
3.12.9 Activation of the V-Live function
Hold the HOLD key pressed and bring the rotary switch from the OFF position to the V position.
The unit emits a double beep then the V and
symbol flashes.
The selected configuration is memorised when the key is released: the V symbol is then ON steady and the symbol
flashes. Proceed in the same way to suppress the V-Live function: the
symbol then goes OFF when the key is
released.
3.12.10 Modification of audio indication threshold in continuity test
Hold the Yellow key pressed and bring the rotary switch from the OFF position to the position
. The unit beeps, the
W
and symbols go ON along with the threshold value (40.0 by default). Adjustment is then possible
from 1
W
to 40
W
by successive pressings on the yellow key (short pressure: progression of 1
W
by 1
W
; press and hol
hold: progression of 10
W
by 10
W
). Once the value is chosen, actuate the rotary switch to memorise.
3.12.11 Default configuration of the unit
Hold the Yellow key pressed and bring the rotary switch from the OFF position to A position.
The unit emits a double beep then all the segments of the digital display and the
symbol flash.
The default configuration is memorised when the key is released: the display no longer flashes and the
symbol disappears.
The default configuration is:
- Audio identification threshold: 40
W
- Auto cut-off: with
- V-Live function: none
3.12.12 Date of last calibration carried out on unit
Hold the MIN/MAX key pressed and bring the rotary switch from the OFF position to the V position.
The unit emits a beep, then the calibration date of the unit is displayed in the form week - year (SS.AA) as long as the
MIN/MAX key is pressed.
9