3.12.3 Automatic compensation for cable resistance

This compensation is obtained by holding down the HOLD key when the continuity test or measurement resistance

function is selected.

When the key is released, when the display shows zero, the correction value is memorised in the clamp.

If the value measured is higher than 2 

W

)

  this  correction is stopped and the memorised correction value is reset to

zero.

This correction is prohibited in MIN/MAX mode.

3.12.4 Automatic compensation of current measurement zero

This compensation is obtained by holding down the HOLD key when the current measurement function is selected.

When the key is released, when the display shows zero, the correction value is memorised in the clamp.

 If the value measured is higher than 6 A, this correction is stopped and the memorised correction value is reset to zero.
This correction is prohibited in MIN/MAX mode.

3.12.5 Manual selection of AC, DC or AC+DC mode

By default the clamp switches to AC or DC mode automatically (AC or DC symbol flashing) for the V and A functions.

Short successive pressures on the yellow key are used to select manually AC measurement, DC measurement of compound

signals (AC+DC) and return to automatic mode. When the mode is selected manually, the AC, DC or AC+DC symbol is

fixed. The selection of this manual mode is impossible in MIN/MAX or HOLD mode.

3.12.6 3.12.6 Selections possible in continuity function

By default the clamp is in continuity function  ( 

) for the position corresponding to the rotary switch. Successive presses

on the yellow key enable selection of the resistance measurement (

W

), then the semi-conductor test function (       )

and return to the continuity function (  

)

3.12.7 Selection OF INRUSH function

This is done in function A (AC) by pressing first on the MIN/MAX key then on the yellow key.

Consultation of the values corresponding to this function is possible by pressing first on the HOLD key then by short

successive presses on the yellow key (see § 3.7).

Quitting this function is possible by short presses on the MIN/MAX key.

3.12.8 Suppression of automatic stop

Hold the HOLD key pressed and bring the rotary switch from the OFF position to the position The unit emits a double

beep then the   symbol flashes. The selected configuration is memorised when the key is released: the

 symbol remains lit continuously. The unit returns to automatic stop mode when the switch goes to the OFF position 

.

L’apparecchio emette un duplice bip quindi il simbolo lampeggia.

La configurazione scelta viene memorizzata quando si rilascia il tasto : il simbolo  è acceso fisso.

L’apparecchio ritorna in modo arresto automatico quando il commutatore passa sulla posizione OFF.

3.12.9 Activation of the V-Live function

Hold the HOLD key pressed and bring the rotary switch from the OFF position to the V position.
The unit emits a double beep then the V and    

 symbol flashes.

The selected configuration is memorised when the key is released: the V symbol  is then ON steady and the  symbol
flashes. Proceed in the same way to suppress the V-Live function: the  

  symbol then goes OFF when the key is

released.

3.12.10 Modification of audio indication threshold in continuity test

Hold the Yellow key pressed and bring the rotary switch from the OFF position to the position

 . The unit beeps, the 

W

 and symbols go ON along with the threshold value (40.0 by default). Adjustment is then possible

    from 1 

W

 to 40 

W

 by successive pressings on the yellow key (short pressure: progression of 1 

W

 by 1 

W

; press and hol

    hold: progression of 10 

W

 by 10 

W

). Once the value is chosen, actuate the rotary switch to memorise.

3.12.11 Default configuration of the unit

Hold the Yellow key pressed and bring the rotary switch from the OFF position to A position.
The unit emits a double beep then all the segments of the digital display and the  

 symbol flash.

The default configuration is memorised when the key is released: the display no longer flashes and the

 symbol disappears.

The default configuration is:

-  Audio identification threshold: 40 

W

-  Auto cut-off: with

-  V-Live function: none

3.12.12  Date of last calibration carried out on unit

Hold the MIN/MAX key pressed and bring the rotary switch from the OFF position to the V position.

The unit emits a beep, then the calibration date of the unit is displayed in the form “week - year” (SS.AA) as long as the

MIN/MAX key is pressed.

9

Содержание VIP CLAMP MINI Series

Страница 1: ...e clamp box is not properly closed Before any measurement check the proper positioning of the cables and switch When measuring current check for proper alignement of the conductor in relation to the m...

Страница 2: ...ement zer 9 3 12 5 Manual selection of AC DC or AC DC mode 9 3 12 6 Selections possible in continuity function 9 3 12 7 Selectione OF INRUSH function 9 3 12 8 Suppression of automatic stop 9 3 12 9 Ac...

Страница 3: ...ty measurement by use of the yellow resistance key and semi conductor test A Measurement of DC and AC voltages rms value PF Measurement of active reactive and apparent var W powers and power factor in...

Страница 4: ...tal display The liquid crystal display includes the digital display of the measured values the related units and symbols Digital display 4 digits 9999 points 3 decimal points and signs DC and peak mea...

Страница 5: ...V 3 Connect the assembly to the voltage source to be measured making sure if possible that this voltage does not exceed the maximum acceptable limits see table below Range switching and AC DC selecti...

Страница 6: ...asuring range Accuracy 1 Resolution Open circuit voltage Protection 0 0 399 9W 1 L 2 pto 0 1 W 300 A Measuring current 3 2 V 500V AC o 750V DC or peak MIN MAX Mode Accuracy same as previous table 0 2...

Страница 7: ...o in DC Repeatability of the measurement after several consecutive closings of the clamp 0 3 typical MIN MAX Mode Accuracy same as previous table 0 2 L Capture time 100 ms typ PEAK Mode Accuracy same...

Страница 8: ...3f arrow on the jaws must be directed in the direction of power circulation from the source to the load In which case the sign corresponds to power consumed by the load the sign corresponds to power s...

Страница 9: ...he frequency measurement is performed on the voltage signal 1 Perform a short pressing on the Hz key the display shows the frequency of the measured signal 2 Pressing again returns to the previously d...

Страница 10: ...essing unlocks the display 3 12 2 Preselection of MIN MAX mode A short pressing on the HOLD key then on the MIN MAX key preselects the MIN MAX mode A second pressing on the HOLD key makes the MIN MAX...

Страница 11: ...12 8 Suppression of automatic stop Hold the HOLD key pressed and bring the rotary switch from the OFF position to the position The unit emits a double beep then the symbol flashes The selected configu...

Страница 12: ...x 10 s measurements 6 Battery charge indicator Flashing charge life 1 h Fixed Change battery Automatic OFF after 10 minutes with no action on the switch or the keys reactivation by switching from the...

Страница 13: ...V V I I P P C C L L A A M M P P M M I I N N I I 4 6 Variations in operating range 11...

Страница 14: ...t the top of the battery cover rear of the clamp and push the battery cover upwards Replace the used battery by a 9 V battery type LF22 observe the polarities Install the battery in its housing then i...

Страница 15: ...V V I I P P C C L L A A M M P P M M I I N N I I APPENDIX APPENDIX 13...

Отзывы: