EnergyPro Datalogger
Logger Operation
If the
EnergyPro
cannot identify the connection type or compensate for a
misconnection then the voltage lead and CT icons will be rearranged according to the
EnergyPro
determinations but there will be no configuration description highlighted.
On the main screen the “running” icon will be replaced by a “connection” icon and
start recording will be inhibited.
Pressing F1, F2 or F3 in the connection screen will override the automatic
configuration function and allow start recording. An example when this might be
done is when measuring three different single phase loads, V1 would be connected to
the voltage and the three CTs would be connected to three different loads all on the
same phase. The automatic configuration function would show the three CTs
connected to the same phase but would not allow recording to start as the 1V3I
configuration is reserved for three phase systems. Pressing F2 will force the 1V3I
configuration and allow recording to start.
Connection Icons:
normal voltage lead icon
reverse voltage lead icon
normal clamp CT icon
reverse clamp CT icon
normal Flex CT icon
reverse Flex CT icon
If the voltage or current icon disappears from the screen then the voltage or current is
too low. If the voltage or current icon is flashing then the voltage or current signal is
too high.