background image

Once  all  the  batch  settings  have  been  created,  press 

Batch/New 

Batch/Open Batch 1 - 

where ‘Batch 1'

 

may now be a re-named as 

required.

Alpha numeric naming batches

The user can rename the batch to a more appropriate name at any 
time.
When  a  new  batch  is  created,  the  gauge  automatically  names  it 
‘Batch n’ - where ‘n’ is the next unused batch number; e.g. Batch 1, 
Batch 2, Batch 3, etc.

To rename a batch either:

Ÿ

Press 

Batch/New Batch/Rename Batch 1

; or

Ÿ

Press 

Batch/Edit Batch/Rename Batch

 and select the Batch to 

be renamed from the list of existing batches

The current name of the batch is in white 
in the middle of the display with a yellow 
‘cursor’ icon to the right of the name.

Using  the  left  arrow,  right  arrow  and 
Select softkeys, the user can rename the 
batch as required.
The yellow, raised character or function 
is  the  item  that  will  be  ‘chosen’  by 
pressing the Select softkey.

Pressing  the 

Escape 

softkey  at  any  time  cancels  the  whole 

operation without making any changes to the original name.

To  store  the  re-named  ‘name’  highlight  OK  and  press  the 
Select softkey.

www.elcometer.com

22

R

en

Содержание E130-S

Страница 1: ...Operating Instructions The New Elcometer 130 Salt Contamination Meter...

Страница 2: ...r in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off andon theuserisencouragedtotrytoc...

Страница 3: ...CHING T 21 9 DOWNLOADING DATA USING ELCOMASTER 2 0 28 10 ELCOMASTER MOBILE 48 11 CLOUD COMPUTING WITH ELCOMASTER T 50 12 GETTING THE MOST OUT OF ELCOMASTER 2 0 51 13 CHANGING THE WELCOME SCREEN 53 14...

Страница 4: ...Contamination Meter quickly and accurately measures the level of soluble salts on surfaces The gauge is available in two models Model S andT This manual describes the operation of both Elcometer 130...

Страница 5: ...c Feet 12 Battery Door 13 Tripod Mount Point 4 x AA Batteries USB Cable ElcoMaster 2 0 Software Model T Test Certificate Calibration Certificate dependent on version ordered User Guide Elcometer 130 S...

Страница 6: ...ol orange batteries at full capacity Empty symbol red flashing batteries at lowest sustainable level To maximise battery life ensure that Auto Screen Brightness is selected or the screen brightness is...

Страница 7: ...reading screen Selecting Auto will enable the gauge to automatically select the appropriate brightness for the surrounding light conditions using the gauge s internal light sensor Screen brightness c...

Страница 8: ...ter paper size detected or selected assuming that the paper is positioned on the measurement electrodes correctly and if manually set up the correct papersizeisused Nomanualcalculationsarerequiredbyth...

Страница 9: ...equivalent to 2 2 g cm or 119 ppm can be offset using the Calibration Offset feature T see page 7 6 Using tweezers remove a filter paper from the pack 4 For half size paper fold the standard paper in...

Страница 10: ...e measurement electrodes correctly and if manually set up the correct paper size is used No manual calculations arerequiredbytheuser 12 Lift the lid and remove the filter paper 13 Place the filter pap...

Страница 11: ...ose the lid ensuring that the magnetic catch is fully engaged 9 The reading will automatically be displayed on screen and stored into memory 10 Lift the lid and remove the filter paper 11 Place the fi...

Страница 12: ...auge at 0 4 5 and 20 g cm nominal values If the gauge tile readings are outside the stated accuracy of the gauge re calibration is recommended contact Elcometer or your local supplier for further info...

Страница 13: ...when timer is enabled ST k Reading Value ST l User Selectable Statistics up to 8 can be displayed T m Batch Name displayed when in batching T Date Time displayed when not in batching if enabled ST n...

Страница 14: ...ed on via Menu Setup Screen Settings Screen Brightness Auto Gauge Welcome Screen ST Users can adjust the start up screen on their Elcometer 130 gauges via the ElcoMaster 2 0 software see page 53 for m...

Страница 15: ...ter 130 is connected via USB to any computer with an internet connection and with ElcoMaster 2 0 installed ElcoMaster 2 0 highlights the availability of a firmware update identifies to the user the ne...

Страница 16: ...start stop and reset the timer as required Calibration Offset T Soluble salts should be measured using high purity water Non pure water with a maximum conductivity of 237 5 S cm equivalent to 2 2 g cm...

Страница 17: ...djusts the reading accordingly To activate press Menu Setup Temperature Comp The Elcometer 130 Model S does not have automatic temperature compensation and therefore the actual level of soluble salts...

Страница 18: ...eeds the limit the gauge notifies the user by turning the reading and limit icon red flashing the red LED and sounding the alarm beep For more information see Limits on page 20 On Screen Statistics T...

Страница 19: ...y selecting Display Readings Run Chart the user can view a line trend graph of the last 20 measurements The graph is updated automatically when each reading is taken Up to four horizontal axes are dis...

Страница 20: ...the batch name when the gauge is in Batch mode T When a batch is created re opened or when the gauge is switched on whilst in batch mode and when each reading is taken the time and date details are st...

Страница 21: ...The Red LED flashes The reading is displayed in red The alarm beeps 3 times The limit icon turns red In addition the number of readings above limit statistic is incremented when readings are stored i...

Страница 22: ...readings statistics batch information and a graph of the readings Batch Review Batch Delete a Batch or all batches entirely from the gauge Batch Edit Batch Delete Batch Delete the last reading entirel...

Страница 23: ...ress Batch Edit Batch Rename Batch and select the Batch to be renamed from the list of existing batches The current name of the batch is in white in the middle of the display with a yellow cursor icon...

Страница 24: ...me name Copying an existing batch By selecting Batch Copy Batch the user will find a list of all the batches that have been created together with the number of readings stored in each batch To copy a...

Страница 25: ...use the up and down arrow softkeys and select the appropriate batch After a request for confirmation the gauge will clear the readings from the selected batch and return the gauge to the list of batc...

Страница 26: ...he appropriate batch After a request for confirmation the gauge will delete the selected batch and return the gauge to the list of batches To delete another batch repeat the procedure to return the ga...

Страница 27: ...h Open Existing Batch and select the last batch in the link If the batch is full the gauge will prompt the user to open a new batch before continuing If the batch is partially complete theusercancompl...

Страница 28: ...inthebatchthancan be displayed clearly on a single screen multiple readings will be combined into one bar Should a single reading within the combined bar be above the upper limit the wholebarwillbered...

Страница 29: ...truction book explains how users can download data using ElcoMaster 2 0 Installing ElcoMaster 2 0 To install ElcoMaster 2 0 insert the supplied mini CD into your CD ROM drive The software will automat...

Страница 30: ...button and click Next ElcoMaster 2 0 will automatically select a default destination location for the software to be installed Click the Browse button to select an alternate file location Click Next...

Страница 31: ...ow provides a list of the settings that have been selected Click Install to undertake the installation or Back to review or alter any settings ElcoMaster 2 0 will now install Press Finish to exit Setu...

Страница 32: ...and an expand arrow Pressing the Menu button opens up a drop down list of options Help Provides Elcometer contact details and information about ElcoMaster 2 0 Update Provides users with the ability to...

Страница 33: ...ram chart Alter the colour settings of the user interface The text file separator used when exporting files Select an alternate location to store the database Enable or disable the auto update option...

Страница 34: ...xcel Link and Standard Mode plus Design custom reports View data in standard forms predefined within the Design Report action tab or by Elcometer Design a collection template to guide inspectors where...

Страница 35: ...d and reported using the ElcoMaster 2 0 software These instructions will inform the user how to download data from an Elcometer 130 Salt Contamination Meter All appropriate Elcometer gauges can be dow...

Страница 36: ...ElcoMaster 2 0 is installed onto a computer without Bluetooth capability however the Bluetooth tab is greyed out as displayed in top image Select the appropriate connection method Switch on the Elcom...

Страница 37: ...he Bluetooth PIN Code and press Next ElcoMaster 2 0 will now validate the connection press Finish when complete The gauge is now connected and ElcoMaster 2 0 will return the user to the Download actio...

Страница 38: ...ect a single or multiple spreadsheet option Single all batches selected will be downloaded into a single spreadsheet Multiple each batch will be downloaded into a separate spreadsheet tab Select one o...

Страница 39: ...s Next Select or create a new folder to save the gauge s data to In some cases users may have re used an existing batch name in the gauge The user should now chose either to select Rename Duplicate Fi...

Страница 40: ...also allows users to manually input data as required This instruction book provides a brief summary of some of the features of ElcoMaster 2 0 however users should read the ElcoMaster 2 0 help file for...

Страница 41: ...g the computer mouse s right button over a file orfolderprovidestheuserwiththefollowingfunctions New Allows users to create a new Project create a new Folder import an Image import a PDF Document crea...

Страница 42: ...txt Document create a new batch type Move to Moves the selected items to a different folder Print Prints a report of the selected items PDF Generates a PDF report of the selected items Email Generates...

Страница 43: ...he following RH Relative Humidity Ts SurfaceTemperature Ta AirTemperature Td DewpointTemperature T DeltaT Ts Td Tdb Dry BulbTemperature Twb Wet bulbTemperature SH Specific Humidity Right mouse click o...

Страница 44: ...ng a colour Measurements Displays all the individual readings Users can alter the display order Date Time Relative Time seconds Relative Time minutes and Reading Number by clicking on the column title...

Страница 45: ...e Structure name Structure location Batch Label Examples Inspector name shift number area under consideration strutt number etc Photos Photographs of the area under inspection specific defects etc can...

Страница 46: ...ction method has been set up see pages 34 36 ElcoMaster 2 0 will then link with the selected gauge and when ready display the gauge information within the Live Readings window At any time the user can...

Страница 47: ...h to begin the design To design the collect image sequence simply click and hold each reading circle in turn drag onto the image where readings are to be taken and adjust the size accordingly Once com...

Страница 48: ...Select a gauge from the list Select the relevant connection method ElcoMaster 2 0 will link the selected gauge and when ready display the gauge information within the Live Readings window Once connect...

Страница 49: ...e Profile Gauge Elcometer 319 Dewpoint Meter Elcometer 456 CoatingThickness Gauge Store live readings directly on the phone and save them into batches Download batches straight to a mobile phone or ta...

Страница 50: ...id com elcometer elcomaster and follow the on screen instructions Download collection images created using the Design Collect function in ElcoMaster 2 0 to indicate where and in which order readings...

Страница 51: ...t will have instant access to all inspection data in real time Data can be accessed by authorised users via secure log in from any computer or mobile device anywhere in the world via Bluetooth USB WiF...

Страница 52: ...secure log in on any computer or mobile device anywhere in the world 2 Gauge to PC data transfer into ElcoMaster 2 0 Using Bluetooth or USB ElcoMaster transfers inspection data in seconds archiving d...

Страница 53: ...m sub assembly plants Combine all data from sub Your data your choice your control assembly and final assembly ElcoMaster 2 0 allows you to inspection to generate Project decide which Cloud service pr...

Страница 54: ...er welcome screen Switch on your gauge Select the DownloadActionTab Click the Edit button on the right hand side of the appropriate gauge Select Change Welcome Screen from the drop down menu Select th...

Страница 55: ...ction only Clicking the Update Gauge button will display a list of the Fixes in all the updates that have been released since the gauge s last upgrade Press Next to continue Press Next to commence dow...

Страница 56: ...d icon T Filter Paper Size Auto Selected No Paper Detected ST Filter Paper Size Full ST Filter Paper Size Half ST Filter Paper Size Quarter ST Timer Enabled ST Temperature Compensation On T Calibratio...

Страница 57: ...130 Model S About Gauge Information Contact Legal Menu Setup About Reset Emergency Light Legal Legal Notices Regulatory Filter Paper Size Auto Full Half Quarter Time and Date SetTime Set Date Set For...

Страница 58: ...Edit Batch Copy Batch Review Batch Deleted Reading Display Clear Statistics Readings Only Readings Selected Stats Readings Run Chart Readings Bar Graph Batch Graph Statistics Review Batch Batch 1 Sta...

Страница 59: ...ed or USB Battery Life Approximately 4 000 measurements Dimensions 250 x 145 x 50mm 9 8 x 5 7 x 1 9 Weight 780g 1 72lb 18 SPARES ACCESSORIES Description Part Number Calibration Verification Tiles Set...

Страница 60: ...lcometer com website Consumable items have a 1 year warranty from the date of purchase against manufacturing defects only This gauge incorporates a Liquid Crystal Display LCD If the display reaches a...

Страница 61: ...ification 59 Character Size 13 Conductivity Measuring 10 D Date and Time 19 Delete Readings 21 Dimensions 58 Display 11 Brightness 6 Reading Size 13 Downloading Data via Bluetooth 14 28 via ElcoMaster...

Страница 62: ...ay 11 Run Chart 18 S Setting Limits 20 Soluble Salts Measuring 8 Software ElcoMaster 2 0 28 SSPC Guide 15 59 Standards 59 Statistics 17 Switching On Off 5 T Technical Specifications 58 Temperature Com...

Страница 63: ...ingapore 589472 Tel 65 6462 2822 Fax 65 6462 2860 e mail asia elcometer com www elcometer com ENGLAND Elcometer Limited Edge Lane Manchester M43 6BU Tel 44 0 161 371 6000 Fax 44 0 161 371 6010 e mail...

Отзывы: