User Guide

Elcometer 407L

Dual &Triple Angle  

  

Glossmeters

en

R

Содержание 407L

Страница 1: ...User Guide Elcometer 407L Dual Triple Angle Glossmeters en R...

Страница 2: ...d transmitted transcribed stored in a retrieval system or otherwise or translated into any language in any form or by any means electronic mechanical magnetic optical manual or otherwise without the p...

Страница 3: ...2 On Off Key 3 USB Data Output Socket 4 Calibration Tile Magnetised Elcometer 407L Glossmeter ISO 17025 High Gloss CalibrationTile with Certificate GlossTile Cleaning Cloth Transit Case Mains Charger...

Страница 4: ...press and hold the On Off key for 3 seconds The gauge will switch off automatically after 30 60 90 or 120 seconds of inactivity as set by the User via Menu Setup Instrument Setup This feature can als...

Страница 5: ...PARAMETERS Via Menu Setup Measurement Menu users can select which measurement values are displayed Geometry Display gloss readings at 1 2 or 3 gloss angles Range Select AUTO full scale or BLACK standa...

Страница 6: ...ed in Section 4 3 ress the or navigation button to highlight CALIBRATE followed by to select Press the or navigation button to highlight CONFIRM followed by to select and save the calibration 4 2 CALI...

Страница 7: ...lace the gauge on the substrate and press the navigation button 2 The reading is taken and the measurement data updated The gauge can be set to take automatic readings at 2 5 or 10 second intervals Th...

Страница 8: ...ings are automatically stored in the open batch until a new batch is created Batches are named BATCH xxn by default where xxn is incremented for each new batch created BATCH can be changed to a user d...

Страница 9: ...ailable in Japan 2 0 Software supplied with the gauge and available as a free download at the gauge can transmit readings to a PC for archiving and report generation Data can be transferred via USB 2...

Страница 10: ...in programs such as Notepad Excel or Wordpad User defined names should entered into this document with each name listed on a separate line The batch name length is limited to 8 letters or numbers furt...

Страница 11: ...Setup Data Batch Menu Calibration Menu Measurement Menu Control Menu Instrument Setup Factory Menu About Calibration Calibrate Instrument Auto Calibration ON OFF Calibrate to xxxxx Tile Value 20 xx x...

Страница 12: ...ON Resolution 0 1 GU Repeatability 0 2 GU Reproducibility 0 5 GU Power USB rechargeable in 4 5 hours the Elcometer 407L can only be charged via USB Battery Life 8hrs continuous operation or 10 000 rea...

Страница 13: ...e measurement angles most commonly used for gloss are 20 60 and 85 The most appropriate angle should be selected dependent on the glossiness of the sample surface Using the correct measurement geometr...

Страница 14: ...ce pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radi...