The Elcometer 130 Model S does not have automatic temperature compensation and therefore, the
actual level of soluble salts needs to be calculated manually using the following equation:
2
µg/cm
If using the Elcometer 130 Model T with Automatic Temperature Compensation disabled, the same
equation should be used to correct the gauge reading for temperature variances.
Where:
k
= Corrected Result
c
= Temperature Coefficient (2.0%)
T
= Measurement Temperature °C
T
= Calibration Temperature (25°C)
cal
Example: Measured Value: 12.3
at 28°C
3 GETTING STARTED (continued)
k
=
k
measured
1 +
c
[
T
-
T
]
cal
k
=
12.3
2
11.6
µg/cm
1 +
0.02
[
28
-
25
]
en
R
www.elcometer.com