Manuale di istruzioni/Scheda tecnica

La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.

The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.

La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.

ART. 09/08550-00

PAGINA 5 DI 6

3. Specification:

The instrument shall be re-calibrated at an interval of one year under the 

conditions of 18°C-28°C and relative humidity of less than 75%. 

3.1 Overview

• The instrument is designed to automatically select measurement functions 

and measuring ranges.

• Full range overload protection.

• Allowable max voltage between terminal 

  to be measured and ground: 

600VDC or 600VAC

• Altitude:  

max 2000 mt.

• Display unit:  

LCD

• Max display value:  

6000 numbers

• Polarity indication:   automatic indication. "-" indicates negative polarity

• Over range indication:  

"0L" or "-0L"

• Sampling rate:  

approximately 3 times per second

• Unit display:  

to display functions and electric quantity

• Automatic shut-down time: 

10 minutes

• Power supply:  

1.5V AAA battery ×2

• Battery under-voltage indication:  

LCD display symbol

• Temperature coefficient:  

less than0.1x accuracy/°C

• Working temperature: 

18°C÷28°C

• Storage temperature: 

-10°C÷50°C

3.2 Technical parameters

3.2.1 AC current

Measuring range

Resolution

Accuracy

6A

0.001A

±(2.5% R 8 digits)

60A

0.01A

200A

0.1A

-Minimum input current: 0.01A  AC current

-Max input current: 200A  AC current 

-Frequency range: 45÷65Hz 

3.2.2 DC voltage

Measuring range

Resolution

Accuracy

600V

0.1V

±(0.5% R 3 digits)

-Minimal input voltage: 0.5V DC

-Maximal input voltage: 600V DC

3.2.3 AC voltage

Measuring range

Resolution

Accuracy

600V

0.1V

±(0.8% R 5 digits)

-Minimal input voltage: 1.0VAC

-Maximal input voltage: 600VAC (valid value)

-Frequency range: 45÷65Hz

3.2.4 Frequency

3.2.4.1 Frequency measurement for the current (A level) through clamp head

Measuring range

Resolution

Accuracy

60.0Hz

0.1Hz

±(1.0% R 5 digits)

1000Hz

1Hz

-Measurement range: 40Hz÷1000Hz

-Input signal range: ≥ 0.2A AC current(valid value)

3.2.4.2 Frequency measurement for the current (V level) through clamp head

Measuring range

Resolution

Accuracy

60.0Hz

0.1Hz

±(1.0% R 5 digits)

1000Hz

1Hz

-Measurement range:40Hz÷1000Hz

-Input signal range:≥ 0.2A AC voltage(valid value)

3.2.5 Electric resistance

Measuring range

Resolution

Accuracy

6KΩ

0.001KΩ

±0.8% R3 digits

-Overload protection: 600VDC or AC (valid value)

3.2.6 Line On/Off test

Measuring range

Resolution

Functions

   

1W

If  the  electric  resistance  of 

the line measured is less than 

50W,  the  buzzer  inside  the 

instrument may sound.

-Overload protection: 600VDC or AC (valid value)

4. Operational guidelines

4.1 Reading hold

During the process of measurement, gently touch the key               if you want 

to hold readings, and monitor’s display value will be locked. 

Touch again the key           , the readings hold will be removed.

4.2 Backlight

1. In the process of measurement, if the measurement environment is too 

dark, press the key           for more than 2 seconds to enable backlight 

function.  Then  about  1  minute  later,  the  backlight  function  will  be 

automatically disabled.

2. During this process, press the key             for 2 sec. to disable backlight.

4.3 Auto shutdown

1. If  no  operations  occur  within  10  minutes  after  the  initialization,  the 

instrument will be in the state of dormancy. Auto shutdown at this moment 

can save power consumption. 2 minutes before shutdown, the buzzer will 

sound at an interval of 1 minute.

2. Press any key after auto shutdown to wake the instrument into operation.

3. The  function  of  auto  shutdown  will  be  disabled  if  Hz/NCV  key  is 

pressed while the instrument is initialized.

4.4 Preparation for measurement

1. Press the power key for 2 seconds to initialize the instrument. If the battery 

voltage is low (approximately ≤2.4V), the monitor will display the symbol 

. At this moment, the battery should be replaced. The instrument will 

be shut down if the power key is pressed after the initialization.

2. When the instrument is not used for measurement, it will enter the 

automatic scan status and display the diagram below.

4.5 Non contact voltage detection (NCV)

1. Press  Hz/NVC  key  for  2  seconds  to  enable  NCV  function.  Then  the 

instrument will display 

2. Press  NCV  key  and  move  NCV  sensor  closer  to  the  lead  line  to  be 

measured.  The  instrument  can  detect  whether  the  AC  voltage  of  the 

measured lead line is >90V. When the instrument detects AC voltage, 

the buzzer will sound alarms while backlight flickers.

  

H

/⁕

H

/⁕

H

/⁕

H

/⁕