DIAMAG
Operating Manual
Page 23
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
Complies
Portable and mobile RF
communications
equip-
ment should be used no
closer to any part of the
device, including cables,
than the recommended
separation distance calcu-
lated from the equation
applicable to the frequen-
cy of the transmitter.
Recommended separation
distance:
d = 1.17 √P
d = 1.17 √P 80 MHz to
800 MHz
d = 2.33 √P 800 MHz to
2.5 GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmit-
ter manufacturer and d is
the recommended separa-
tion distance in metres
(m).
Field strengths from fixed
RF transmitters, as de-
termined by an electro-
magnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur
in the vicinity of equip-
ment marked with the fol-
lowing symbol
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
Complies
Note:
Ut is the AC mains voltage prior to application of the test level.
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propaga-
tion is affected by absorption and reflection from structures, objects and
people.