Page 10 of 12
Quick Guide
EC-Link-Software (PAT-Tray, PAT-Stand-16, PAT-Chamber-16)
Release 2.5
6 Settings
The settings of the EC-Link software being specific to the instrument are contained in the file
C:\ProgramData\EC-Link\Settings… A copy of this file can also be found on the installation CD
and may be used to restore the original settings. Within the program, in the
Settings/Calibration dialog, you can only modify the number of decimal digits displayed, the
colours of the individual traces, and the maximum error
E of the data reduction algorithm.
The ‘Reduction Algo max Error’
E has a great impact on the efficiency of the data reduction
and hence on the size of the compressed *_Small.txt log file. In short, the algorithm checks if
a given data point is more than
E off the straight line connecting two adjacent data points. If
so, then the given data point is kept in *_Small.txt, otherwise it is discarded. Note that the
algorithm is applied to all channels. Therefore, if
E is set to zero for any channel, then no
data reduction takes place.
In the Settings/Program dialog, you can select additional columns to be logged into the *.txt
and *_Small.txt log files. The UTC time format is especially useful in case different consecutive
measurements need to be referred to the same time scale.