  Page 10 of 21

 

Release 1.21 

      User Manual

 

ECC-Opto-10

 

 

 

 

4.  Put the separator on top and push 
it down using the loading tweezer.  

The separator must not be more than 
10 mm in diameter. A smaller 
separator may be useful to prevent the 
sample holder from coming into 
contact with the electrolyte. 

 

 

5.  Attach the sample holder and press 
it down. 

 

6.  Insert the assembly into the cell 
base. 

Содержание ECC-Opto-10

Страница 1: ...2022 EL Cell GmbH User Manual Release 1 21 ECC Opto 10 Electrochemical test cell...

Страница 2: ...ies EL Cell GmbH maintains the right to make changes without further notice to products described in this manual to improve reliability function or design EL Cell GmbH does not assume any liability ar...

Страница 3: ...0 Content 1 Product Description 4 2 Technical data 6 3 Intended use 6 4 Safety Precautions 6 5 Assembly 7 6 Connecting the test cell 17 7 Disassembly and Cleaning 17 8 Unpacking 18 9 Components ECC Op...

Страница 4: ...ECC Opto 10 is connected to the battery tester via a 2 mm cell cable with banana plugs It can be used with the PAT Tester x 8 as well as potentiostats and battery testers from third party manufacturer...

Страница 5: ...ECC Opto 10 side by side setup of electrodes Sapphire glass window Glass fiber separator Electrode 1 Inspection area up to 18 mm Electrode 1 pin Electrode 2 Lid sealing foil PE Electrode R pin Lower p...

Страница 6: ...environment It may only be used by trained personnel and only as described in this manual 4 Safety Precautions Use proper safety precautions when using hazardous electrode materials and electrolytes...

Страница 7: ...several different configurations which mainly differ in the type and size of the working electrode used the charging geometry the position of the working and counter electrode relative to each other...

Страница 8: ...al ECC Opto 10 5 1 Lid assembly 1 Turn the lid upside down Insert the foil sealing into the lid 2 Insert the window 3 Insert the window thrust ring 4 Attach the window flange and screw in the screws u...

Страница 9: ...the cell with a side by side sample holder 1 Assemble the lid as described above Then attach the electrode strips to the sample holder 2 Use the loading tweezer to push the piston into the sleeve 3 Pu...

Страница 10: ...h it down using the loading tweezer The separator must not be more than 10 mm in diameter A smaller separator may be useful to prevent the sample holder from coming into contact with the electrolyte 5...

Страница 11: ...s Use only the amount of electrolyte necessary to impregnate the porous material Avoid excess electrolyte Typically around 30 l of electrolyte is suitable Note that electrolyte will be squeezed out of...

Страница 12: ...o 10 11 Doublecheck the proper alignment of the electrode strips Then press down the lid and tighten the three screws firmly with the 2 5 mm hexagon screwdriver 12 The ECC Opto 10 is now ready for tes...

Страница 13: ...face to face sample holder 1 Follow the lid assembly steps Then insert the lower piston into the inner sleeve 2 Use the loading tweezer to push the piston into the sleeve 3 Place the assembly into th...

Страница 14: ...ain to push the cell stack further down 7 Drop the electrolyte onto the separator Use only the amount of electrolyte necessary to impregnate the porous material Avoid excess electrolyte Typically less...

Страница 15: ...Opto 10 9 Push down the electrode stack using the loading tweezer 10 Insert the contact ring 11 Put the holed current collector on top Choose the hole size depending on the magnification used by the...

Страница 16: ...ECC Opto 10 13 Put on the cell lid Make sure that the flat side of the cell base and the straight line on the lid are parallel 14 Finally screw the cover on with the 2 5 mm hexagon screwdriver 15 The...

Страница 17: ...e disassemble the cell in the reverse order of assembly All chemicals used have to be disposed of properly All wetted parts are to be cleaned with water and or other appropriate solvents All parts are...

Страница 18: ...th flat socket fully equipped for use in both 2 electrode and 3 electrode configuration Windows seal 10 pcs ECC1 05 0016 B X Separator 10 0 mm x 0 26 mm GF A 10 pcs ECC1 01 0012 R X PE Sealing foil 10...

Страница 19: ...Page 19 of 21 Release 1 21 User Manual ECC Opto 10 9 Components ECC Opto 10 ECC Opto 10...

Страница 20: ...Page 20 of 21 Release 1 21 User Manual ECC Opto 10 Lid OPTO 10 18 mm assy...

Страница 21: ...his Limited Warranty does not cover defects damage or nonconformity resulting from abuse misuse neglect lack of reasonable care modification or the attachment of improper devices to the goods This Lim...

Отзывы: