19
EN
[English]
CAUTION
• Do not use eKuore Pro® Series while charging the battery
• Do not use the product if present any damage
• Do not use the device over open wound or on infected skin
• Ensure that the chestpiece is tightly curled as it can interfere with the operation of eKuore Pro® Series
• Ensure that the chest piece is screwed on carefully and tightly so as not to damage the membrane or interfere with the device operation.
• Clean the stethoscope from patient to patient.
• In case of electrostatic discharges, the mobile phone/tablet may behave abnormally. In this case, restart the phone/tablet.
• Follow the operating instructions, since misuse can cause leakage currents.
• Do not use the eKuore Pro Series in environments with high interferences or electromagnetism.
• Check the product box before use, verify that it contains all necessary components.
• eKuore Pro Series requires for EMC precautions to be incorporated.
• eKuore Pro Series can be affected by radiofrequency communications.
• The use of accessories other than those indicated by eKuore may result in increased emissions or decreased immunity of the eKuore
Pro Series.
• The eKuore Pro Series is powered by an internal battery for continuous operation.
• Do not use the product if it does not work properly.
• The eKuore Pro Series has no defibrillation protection in place.
• The equipment may be subject to interferences due to electrosurgery or high-frequency electrosurgery
• MR-unsafe! Do not expose the device to a magnetic resonance (MR) environment. The device may present a risk of projectile injury due
to the presence of ferromagnetic materials that can be attracted by the MR magnet core. Thermal injury and burns may occur due to
the metal components of the device that can heat during MR scanning. The device may generate artifacts in the MR image. The device
may not function properly due to the strong magnetic and radiofrequency fields generated by the MR scanner.