EKO RoMiotto MS-711 Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

 

 

 

Tel.: +11 3976-4003  •  3999-7737

[email protected]

www.romiotto.com.br

Содержание RoMiotto MS-711

Страница 1: ...Tel 11 3976 4003 3999 7737 vendas romiotto com br www romiotto com br...

Страница 2: ...ninstallation 17 6 2 Data File Format 22 7 Measurement Software WSDac 24 7 1 Starting Measurement Software W SDac 24 7 2 Before Starting Measurement 25 7 3 Start Measurement 27 7 4 Measurement Data Di...

Страница 3: ...troradiometer WISER Instruction Manual Ver 2 Pg 2 11 Specification 44 11 1 Sensor Specifications 44 11 2 Power Supply Specification 45 11 3 Software 45 11 4 Cables 46 11 5 Dimensions 47 11 6 Accessori...

Страница 4: ...el 31 0 70 3050117 Fax 31 0 70 3840607 North America Region www eko usa com info eko usa com EKO INSTRUMENTS USA Inc 95 South Market Street Suite 300 San Jose CA 95113 USA Tel 1 408 977 7751 Fax 1 408...

Страница 5: ...collection points Disposing of this product correctly will help save valuable resources and prevent any potential negative effects on human health and the environment which could otherwise arise from...

Страница 6: ...e following harmonized standards of other normative documents Harmonized standards EN 61326 1 2006 Class B Emission EN61000 3 2 A1 A2 2009 EN61000 3 3 2008 EN 61326 1 2006 Table 1 Immunity EN61000 4 2...

Страница 7: ...ates is in conformity with the following harmonized standards of other normative documents Harmonized standards EN 61326 1997 A1 A2 Class A Emission EN 61326 1997 A1 A2 Immunity EN61000 4 2 EN61000 4...

Страница 8: ...Do not place these instruments in places where humidity and or dusts exist extreme difference in temperature of over specification range near object which emits strong magnetic energy or electric wav...

Страница 9: ...de the best performance in a wide operating temperature range The rugged optical design of the diffusor and input optics make the MS spectroradiometer series concept superior to any fiber optic spectr...

Страница 10: ...de Measurement Range With the automatic exposure function MS 711 712 can measure a wide range of solar illumination from low to high illuminance in high S N proportion MS 711 can measure from 300nm to...

Страница 11: ...unit Set of Sensor fixing screws 1 MS 711 Bolts M8 x 75 Washers Nuts x2 each MS 712 Bolts M8 x 90 Washers Nuts x2 each Power supply 2 One power supply unit for each MS 711 and MS 712 Sensor cable 2 S...

Страница 12: ...he intensity of the incident light to comply with cosine characteristics 4 Quartz Dome Dome protects the sensor inside from soiling and effects from rain and wind and it enhances the cosine characteri...

Страница 13: ...n 5 6 Supply Power 12V Gray Blue 6 7 Supply Power ground White Yellow Green 7 DC GND 8 Supply Power 12V Black 8 9 Supply Power Ground Pink 9 DC GND 10 EARTH Black 10 MS 712 Terminal block Wire Color W...

Страница 14: ...ch has a full hemispheric field of view without any obstructions such as buildings trees and mountain In practice it might be difficult to find such locations Therefore some practical recommendations...

Страница 15: ...rn hemisphere the connector should be orientated North in the Southern hemisphere the connector should be orientated South 3 Leveling Adjust the sensors in a horizontal position by using the 2 levelin...

Страница 16: ...and strip so that it will not be blown by wind If the sensors are setup on Sun Tracker make sure to have sufficient length of cable for rotation movement Use screw driver to connect the cable to lug t...

Страница 17: ...Control Ground Pink 9 DC GND 10 EARTH Black 10 7 Connect Power Supply Connect the one end of power cable to AC power supply connector at the rear of power supply Then connect the other end of power ca...

Страница 18: ...drive while the Windows desktop screen is displayed 2 The install disk is included in the CD ROM labeled as Spectroradiometer MS 711 The separate software for the measurement software and data analys...

Страница 19: ...select the appropriate folder NOTE If the Operation System for the installing PC is either one of the Windows Vista 7 or 8 and to change the destination folder UAC User Account Control function will b...

Страница 20: ...button to go to next step 7 After a while the PC display goes dark and User Account Control dialog window appears Click Yes button to start the software installation 8 As the installation is complete...

Страница 21: ...The installation wizard will start follow the wizard which steps are in the similar procedure as the measurement software Once the software installation is completed following shortcut icon will be cr...

Страница 22: ...from Installation Disk Just like when the software was installed startup the software to be uninstalled from the CD ROM Proceed according to the messages appear on the window If the software is alread...

Страница 23: ...at data The spectral radiation data of wavelength interval according to the spectroradiometers specification can be saved in CSV file as well as the as well as the spectral radiation data which wavele...

Страница 24: ...value that are calculated out from the data in each calculated intervals The PAR PPFD are calculated from the data between 400 700nm and Illuminance is calculated by multiplying the standard relative...

Страница 25: ...omatically setup by the software and takes and save the spectrum data NOTE When taking measurements by switching multiple units of MS 711 MS 712 on one computer make sure to do the following Restart t...

Страница 26: ...hen the Measurement Settings window will be displayed as shown on Figure 7 3 COM Port Setup User Comment Integration Range Setup Measurement Time Setup Save Folder for Binary Format File Save Checkbox...

Страница 27: ...e the date and time of the measurement such as YYYYMMDDhh WSD YYYY Year MM Month DD Day and hh Time Wavelength Pitch By selecting the Raw the data is saved with the unique wavelength pitch of the inst...

Страница 28: ...ment from the tool bar or click the Start button Figure 7 6 located at the top the Interval Measurement window is displayed as sown on Figure 7 7 Current Time The current time is displayed Instrument...

Страница 29: ...action necessary Retrying measurement The measurement failed but retrying to start measurement Check communication cable and power supply of the instrument Measurement failed Displayed when the measu...

Страница 30: ...omment Displays the User Comment entered in the Measurement Setting Range of Integration Value Displays the spectrum radiation integration value W m2 within the wavelength range which is setup in Inte...

Страница 31: ...wavelength range to be displayed on graph Graph Color Frame Color and Grid Line By clicking the colored area the color selection dialog window will be displayed and the graph and frame color can be s...

Страница 32: ...Go to File Print from the tool bar or click the Print button Figure 7 12 at the top to print the graph Also to see the print preview condition go to File Print Preview from the tool bar Printer can b...

Страница 33: ...sis Software WSDisp WSDisp exe is data analysis software and it has main function to view the measured data WSD and save as CSV format files 8 1 Starting Data Analysis Software WSDisp exe As the WSDis...

Страница 34: ...op of the window Fig 8 2 When the file selecting dialog window appears select a measurement data file extension WSD file and click Open button to show the data 2 Displaying Data List After the measure...

Страница 35: ...king this button Double clicking the displayed graph can also switch to the Tiled Data window Measured Date Time The date and time of the displayed measurement taken User Comment Displays the User Com...

Страница 36: ...left clicking and moving the mouse the values of each intersection position are shown at the right bottom of the graph 5 Displaying Multiple Selected Data When selecting multiple data in a sequential...

Страница 37: ...than 10 data are selected the colors are repeated Graph colors can be freely changed by opening the WSDisp_V3 0 x x ini file as text editor and editing each RGB values of the COLOR TABLE Also by addi...

Страница 38: ...0 CSV Data Waveform Specify the wavelength pitch for when the Saving Measurement Value in CSV Format By selecting the Raw the data is saved with the unique wavelength pitch of the instrument itself By...

Страница 39: ...ne By clicking the colored area the color selection dialog window will be displayed and the graph and frame color can be specified as desired The grid line of the graph can be shown or not shown by ch...

Страница 40: ...as CSV Format Save the data of PAR W m 2 PPFD umol m 2 s ILLUMINANCE lx and integral W m 2 per each measured time of the selected measurement data in CSV format Save Measurement Data Save the measure...

Страница 41: ...om the tool bar or click the Print button Fig 8 16 at the top to print the graph Also to see the print preview condition go to File Print Preview from the tool bar Printer can be setup from File Print...

Страница 42: ...1 MS 712 sensor part consist of components such as a dome a diffuser a shutter to measure the dark current a grating with diode array to disperse the light Peltier element to control temperature and p...

Страница 43: ...ensor and communication cables making sure they are connected properly and they do not have any excess tensions A disconnected cable may cause abnormal measurement values and operation mistakes Damage...

Страница 44: ...nties contributed by the standard lamp calibration set up and spectroradiometer performance properties 10 3 Troubleshooting Check the following items in case of trouble with the instrument If any ques...

Страница 45: ...Power supply 12VDC Sensor 12VDC Ventilator 100VAC 50 60Hz 12VDC 5VDC Sensor 12VDC 5VDC Ventilator 100VAC 50 60Hz Maximum Power Consumption 50W 66W 65W 81W Dimension Including cover dome 220 197 H mm 2...

Страница 46: ...umidity 0 90 RH No condensation Dimension 320 W x 240 D x 180 H mm 200 W x 120 D x 160 H mm Weight 1kg 1kg 11 3 Software Table 11 4 Software Specifications Details Software Versions Measurement Softwa...

Страница 47: ...n Manual Ver 2 Pg 46 11 4 Cables Table 11 5 Cable Specifications Cables Details Sensor Cable Cable Type Diameter Outer Diameter Cable end Cable Length MVVS 0 5mm 2 x 12pins 10 4mm Y Terminal Model Num...

Страница 48: ...Grating Spectroradiometer WISER Instruction Manual Ver 2 Pg 47 11 5 Dimensions 1 Sensor Part Figure 11 2 Sensor Part Dimensions Left MS 711 Right MS 712 Tel 11 3976 4003 3999 7737 vendas romiotto com...

Страница 49: ...EKO INSTRUMENTS CO LTD Grating Spectroradiometer WISER Instruction Manual Ver 2 Pg 48 2 Power Supply Figure 11 4 MS 712 Power Supply Dimensions Figure 11 3 MS 711 Power Supply Dimensions...

Страница 50: ...2G Double Arm Sun Tracker STR 32G Heavy Duty Sun Tracker STR 32G DNI Measurement Mounting Plate STR 32G Mounting Plate for MS 710 711 Secondary Side Collimation Tube for MS 711 Collimation Tube for MS...

Страница 51: ......