EKO INSTRUMENTS CO., LTD. Ventilation Unit MV-01 Instruction Manual V2
Pg. 15
6. Maintenance & Troubleshooting
6-1. Maintenance
It is recommended to check and do the following:
Table 6-1. Maintenance Items
Items
Frequency
How To
Effects
Visual Check
Monthly
Check for any cracks and scratches
on Fan and Heater Units.
Using damaged parts may lead to
electrical shock or performance
degradation. Replace the damaged
parts. For replacing the parts, please
contact your nearest distributor or
EKO directly.
Check Spirit Level
Weekly
Check the pyranometer spirit level;
readjust
when
necessary.
(applicable only when installed in
horizontal level)
Cause directional response error
Check Cable
Monthly
Check
for
any
damages,
disconnection of cables, and loose
connector.
Damaged cables and/or loose
connectors may lead to malfunction.
Check Installation
Monthly
Check for loose bolts, any damages
to the installation base/plate.
Products may drop due to loosen
bolts and may get damaged.
Check Fan
Operation
Monthly
Check for any obstacles like trash,
plants, and bugs are stuck in the fan;
check for any noise from the fan.
Continuing use of fan with things
stuck in fan may lead to damage
and/or cause fire.
Replace Fan
Every
2 years
Replace the fan. For replacing the
fan, please contact your nearest
distributor or EKO directly.
Continuing use of fan after its lifetime
may lead to fan malfunction,
decrease in rotation frequency, and
operation stop.
Clean Air Filter
(Optional)
Monthly
Clean the air filter
Ventilation capacity will be lower
when the air filter gets soiled.
Replace Air Filter
(Optional)
Within
6 months
Replace the air filter within 6 months.
Ventilation capacity will be lower
when the air filter gets soiled.