Chapter 3 Controlling Monitor Quality
Chapter 3 Controlling Monitor Quality
The monitor includes the Integrated Front Sensor, which enables you to perform calibration and Grayscale
Check independently.
Note
•
Calibration can be performed periodically by using the schedule function of RadiCS LE that is attached to the
monitor. For more information about the setup procedure, refer to the RadiCS LE User’s Manual (stored on the
CD-ROM).
•
RadiCS enables you to configure the schedule settings for Grayscale Check as well as for calibration. For more
information about the setup procedure, refer to the RadiCS User’s Manual.
•
If you want to perform high-level quality control in compliance with medical standards / guidelines, use optional
quality control software kit "RadiCS UX1".
•
To adjust measurement results of the Integrated Front Sensor to those of an EIZO external sensor (UX1 or UX2
sensor), perform correlation between the Integrated Front Sensor and the external sensor using RadiCS / RadiCS
LE. Periodical correlation allows you to maintain the measurement accuracy of the Integrated Front Sensor at a
level equivalent to that of the external sensor.
3-1. Using the SelfCalibration Function
Execute SelfCalibration and view the latest results.
●
Executing SelfCalibration
Procedure
1. Select “RadiCS SelfQC” from the Setting menu, then select .
2. Select “SelfCalibration” for “RadiCS SelfQC”, then select .
3. Select “Execute”, then .
SelfCalibration is executed.
Attention
•
If the power to the PC is turned off or the input signal switches while SelfCalibration is running, the execution
will be canceled.
•
After SelfCalibration is complete, perform Grayscale Check at the same room temperature and illuminance
as those in the actual working environment.
Note
•
The following items can be set by using RadiCS / RadiCS LE. For more information about the setup
procedure, refer to the RadiCS / RadiCS LE User’s Manual.
- Calibration target
- SelfCalibration schedule
- If you have configured the schedule using RadiCS / RadiCS LE, the next execution schedule is displayed in
the advanced menu.
•
The calibration target value can be confirmed by selecting the target mode in "CAL Switch mode" of the
Setting menu.