28
User’s Guide
Electromagnetic Compatibility
Like other electrical medical equipment, the
EX210/EX240W
requires special
precautions to ensure electromagnetic compatibility with other electrical medical devices.
To ensure electromagnetic compatibility (EMC), the EX210/EX240W must be installed and
operated according to the EMC information provided in this manual.
Note
The EX210/EX240W has been designed and tested to comply with IEC
60601-1-2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the EX210/ EX240W����
Warning
Do not use cables or accessories other than those provided
with the EX210/EX240W, as this may result in increased electromagnetic
emissions or decreased immunity to such emissions��
Warning If the EX210/EX240W is used adjacent to or stacked with
other equipment, observe and verify normal operation of the EX210/EX240W
in the configuration in which it will be used prior to using it in a surgical pro-
cedure�� Consult the tables below for guidance in placing the EX210/EX240W��
The Model EX210/EX240W is intended for use in the electromagnetic environment specified below.
The customer or the user of EX210/EX240W should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The Model EX210/EX240W uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
RF emissions CISPR 11
Class B
EX210/EX240W is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes
Harmonic emissions
IEC61000-3-2
Class D
Voltage Fluctuations
IEC61000-3-3
Complies
Manufacturer’s declaration - electromagnetic emission
The Model EX210/EX240W is intended for use in the electromagnetic environment specified below.
The customer or the user of EX210/EX240W should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient
/burst IEC61000-4-4
2 kV for power supply lines
1 kV for input / output lines
2 kV for power supply lines
1 kV for input / output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Manufacturer’s declaration - electromagnetic immunity