background image

28

Appendix

Electromagnetic immunity

The EX3241 has been tested at the following compliance levels according to the testing requirements for professional 

healthcare facility environments defined in IEC / EN60601-1-2.

Customers and users of the EX3241 must ensure that the EX3241 is used in the following environments:

Immunity test

Test level for 

professional  

healthcare facility  

environments 

Compliance level

Electromagnetic environment - 

Guidance

Portable and mobile RF communications equip-

ment should be used no closer to any part of the 

EX3241, including cables, than the recommended 

separation distance calculated from the

equation applicable to the frequency of the trans-

mitter.

Recommended separation distance

Conducted distur-

bances induced by 

RF fields

IEC / EN61000-4-6

3 Vrms 

150 kHz - 80 MHz

3 Vrms

d = 1.2√P

6 Vrms 

ISM bands between  

150 kHz and 80 MHz

6 Vrms

Radiated RF fields

IEC / EN61000-4-3

3 V/m 

80 MHz - 2.7 GHz

3 V/m

d = 1.2√P, 80 MHz - 800 MHz

d = 2.3√P, 800 MHz - 2.7 GHz

Where “P” is the maximum output power rating of 

the transmitter in watts (W) according to the trans-

mitter manufacturer and “d” is the recommended 

separation distance in meters (m).

Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey 

a)

should be less than the compliance level in each 

frequency range 

b)

.

Interference may occur in the vicinity of equipment 

marked with the following symbol.

Note 1

U

T

 is the a.c. mains voltage prior to application of the test level.

Note 2

At 80 MHz and 800 MHz, the higher frequency range applies.

Note 3

Guidelines regarding conducted disturbances induced by RF fields or radiated RF fields may not apply in all situa

-

tions. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Note 4

The ISM bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz, 13.553 MHz to 13.567 MHz, 26.957 

MHz to 27.283 MHz, and 40.66 MHz to 40.70 MHz.

a)

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 

assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be consid

-

ered. If the measured field strength in the location in which the EX3241 is used exceeds the applicable RF compliance 

level above, the EX3241 should be observed to verify normal operation. If abnormal performance is observed, additional 

measures may be necessary, such as reorienting or relocating the EX3241.

b)

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Содержание CuratOR EX3241

Страница 1: ...llation Manual separate volume carefully to familiarize yourself with safe and effective usage Please retain this manual for future reference For monitor adjustment and settings refer to the Installat...

Страница 2: ...unit This product has been adjusted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications N...

Страница 3: ...s Symbols on the unit Symbol This symbol indicates Power Switch Press to turn the monitor s power off Power Switch Press to turn the monitor s power on Direct current Alerting to electrical hazard CAU...

Страница 4: ...t falls disconnect the power immediately and ask your local EIZO representative for advice Do not continue using a damaged unit Using a damaged unit may result in fire or electric shock Use the unit i...

Страница 5: ...result in fire or electric shock Handle the power cord and AC adapter with care Handle the power cord and AC adapter with care Do not place heavy objects on pull or tie the power cord or the AC adapte...

Страница 6: ...e electric shock or equipment damage Do not touch the plug or AC adapter with wet hands Doing so may result in electrical shock Use an easily accessible power outlet This is to facilitate disconnectin...

Страница 7: ...aning the unit while it is plugged into a power outlet may result in electric shock If you plan to leave the unit unused for an extended period of time disconnect the power plug from the wall socket a...

Страница 8: ...EIZO representative The screen may have defective pixels or a small number of light dots on the screen This is due to the characteristics of the panel itself and is not a malfunction of the product D...

Страница 9: ...e Welpas Glutaral Sterihyde Glutaral Cidex Plus28 Ammonia Ammonia water Hydrogen peroxide Hydrogen peroxide solution Alkyldiaminoethylglycine hydrochloride Satenidin solution Benzalkonium chloride Zal...

Страница 10: ...2 1 Before Installing the Product 15 Installation Requirements 15 2 2 Installing the Product 16 2 3 Connecting the Power Cord 18 2 4 Connecting the Cables 19 2 5 Installing the Cable Cover 20 2 6 Tur...

Страница 11: ...ith DisplayPort 1 2 SST Single Stream Transport UHD HDMI signal 1 input Compatible with UHD Other functions S R S C Smart Resolution with Sparse Coding function Reduces blur on input images allowing f...

Страница 12: ...AHM250PS48T Cable cover with screws Monitor attachment screw M4 12 x 4 M6 15 x 4 User Manual CD Installation Manual Outline Dimensions Instructions for Use 1 3 Controls and Functions AC adapter 2 3 1...

Страница 13: ...cutes the function assigned to this button Select items in the menu screen 5 F2 button 6 F3 button 7 F4 button 8 F5 ENTER button Executes the function assigned to this button Select items in the menu...

Страница 14: ...G HD SDI 15 SDI 1 2 output terminals BNC type The signal into the SDI input terminal is output as is SDI 1 is compatible with 12G 6G 3G HD SDI and SDI 2 is compatible with 3G HD SDI 16 DVI D 1 2 input...

Страница 15: ...thorough operational testing of the system cables arms etc in the environment where the product will be used Installation Requirements When installing the monitor ensure that there is adequate space a...

Страница 16: ...nd attach it to achieve the following tilt angles of the monitor Up 45 down 45 Connect the cables after attaching an arm or stand The monitor arm and stand are heavy Dropping them may result in injury...

Страница 17: ...ickness of installation fixture 2 mm 4 5 mm M4 screw M6 screw Installation fixture Screw supplied with the monitor Cabinet Using commercially available screws Installation fixture Commercially availab...

Страница 18: ...hape with the port shape rotate the lock ring clockwise and fix it securely Note If the lock ring is stiff and does not rotate push it further into the monitor and try again to rotate it DC 12G SDI IN...

Страница 19: ...DMI cable is held securely Attention Do not use damaged cables Do not connect or disconnect the signal cable while the monitor is turned on The SDI terminal DVI D terminal DisplayPort terminal and HDM...

Страница 20: ...are not pinched between the cable cover and the monitor Securely tighten the screws in the two locations Screw tightening torque 0 4 N m to 0 7 N m Tool required Phillips screwdriver 2 Avoid subjecti...

Страница 21: ...ctly even though the monitor is functioning properly This message appears when no signal is input Example The message shown on the left may appear because some devices to be connected do not output th...

Страница 22: ...178 178 Brightness typical 700 cd m2 Response Time typical 18 ms black white black Contrast Ratio typical 1000 1 Video Signals Input Terminals SDI 1 BNC 12G 6G 3G HD SDI SDI 2 BNC 3G HD SDI DVI DVI D...

Страница 23: ...ironmental Requirements Temperature 20 C to 60 C 4 F to 140 F Humidity 10 to 90 R H no condensation Air Pressure 540 hPa to 1060 hPa AC adapter Power Input 100 240 VAC 10 50 60 Hz 3 0 A Maximum Power...

Страница 24: ...20 x 1080 50p 56 250 50 000 1920 x 1080 59p 67 433 59 940 1920 x 1080 60p 67 500 60 000 640 x 480 60 31 469 59 940 800 x 600 60 37 879 60 317 1024 x 768 60 48 363 60 004 1280 x 800 60 49 702 59 810 12...

Страница 25: ...in malfunction of the monitor Install the equipment in a controlled environment in which such effects are avoided Classification of Equipment Electric shock protection type Class I EMC class EN60601 1...

Страница 26: ...necessary the equipment or system should be observed to check for normal operation in the configuration in which it will be used When using portable RF communication equipment keep it 30 cm 12 inches...

Страница 27: ...onments Compliance level Electromagnetic environment Guidance Electrostatic dis charge ESD IEC EN61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors...

Страница 28: ...survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage pri...

Страница 29: ...5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 9 5500 5785 a For some services only the uplink frequencies are included b Carrier waves are modulated using a 50 duty cycle square wave signa...

Страница 30: ...and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harm...

Страница 31: ...ZO Corporation All rights reserved 1st Edition February 2019 Printed in Japan 00N0N101A1 IFU EX3241 6 www eizoglobal com 8 5B 153 Shimokashiwano Hakusan Ishikawa 924 8566 Japan Siemensallee 84 76187 K...

Отзывы: