46
Chapter6Troubleshooting
6-4. Built-in calibration sensor and SelfCalibration
problems
Problem
Possible cause and remedy
1. The built-in calibration sensor does not
come out/stays out.
• Turnoffthemainpower,andthenturnitonagainafew
minuteslater.
2. SelfCalibration cannot be executed.
• Checkwhetherthedateandtimearesetcorrectly(see
“Settingmonitordateandtime”(page36)
).
• Checkwhetherthemodeisset(see
whichSelfCalibrationisexecuted.”(page38)
).
• Checkwhetherthecalibrationtargetsaresetcorrectly(see
“SettingAdjustmentTarget”(page39)
).
• SettheSelfCalibrationtargetusingColorNavigator.
3. SelfCalibration failure
• Refertotheerrorcodetable.Ifanerrorcodethatdoes
notappearintheerrorcodetableisdisplayed,contact
yourlocalEIZOrepresentative.
Error Code Table
Error Code
Description
0014
• ExecuteSelfCalibrationagain.
• AdjustusingColorNavigator.
0034
• Thebuilt-incalibrationsensormaynothavecomeoutduringcalibration,orlightmay
haveenteredthesensorduringmeasurement.
• Turnoffthemainpower,waitforafewminutesbeforeturningthepowerbackon,and
thenexecuteSelfCalibrationagain.
0050
• ExecuteSelfCalibrationagain.
0061
• Thebuilt-incalibrationsensormaynothaveworkedcorrectly.
• Checkwhetherthereisanyforeignobjectnearthebuilt-incalibrationsensor.
• ExecuteSelfCalibrationagain.